Document Detail


Evaluation of performance of two high-frequency oscillatory ventilators using a model lung with a position sensor.
MedLine Citation:
PMID:  20949287     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
PURPOSE: High-frequency oscillatory ventilation (HFOV) is thought to protect the lungs of acute respiratory distress syndrome (ARDS) patients. The performance and mechanical characteristics of high-frequency oscillatory ventilators, especially with regard to delivering appropriate tidal volume (V (T)) to compromised lungs, might affect the outcome of patients. We evaluated the performance of two such ventilators using a model lung with a position sensor.
METHODS: We tested the Metran R100 and SensorMedics 3100B. V (T) was measured using the model lung with the compliance set at 20 or 50 ml/cmH(2)O and the resistance at 0 or 20 cmH(2)O/l/s. Oscillator frequency was set at 5, 7, and 9 Hz, and amplitude was set at 25%, 50%, 75%, and 100% (100% being maximum amplitude available at each setting configuration).
RESULTS: At each model lung setting, R100 delivered greater V (T) at 5 Hz. V (T) differences between the ventilators decreased as frequency increased and were negligible at 9 Hz. At each model lung setting and frequency, as amplitude increased from 25% to 100%, V (T) increased proportionally more with R100. With an I:E ratio of 1:1, 3100B delivered greater V (T) than with 1:2.
CONCLUSION: Because it is able to deliver comparably greater V (T), R100 may be a better choice for HFOV in critical ARDS patients. Better proportionality may be a result of more effective amplitude titration for adjusting PaCO(2) during oscillation.
Authors:
Naoya Iguchi; Osamu Hirao; Akinori Uchiyama; Takashi Mashimo; Masaji Nishimura; Yuji Fujino
Related Documents :
16018487 - Equal sensation curves for whole-body vibration expressed as a function of driving force.
10485567 - Application of radiation force in noncontact measurement of the elastic parameters.
11008807 - The vibration behavior of railway track at high frequencies under multiple preloads and...
1754307 - Ventilatory sensitivities of peripheral and central chemoreceptors of young piglets to ...
21667267 - The oscar, astronotus ocellatus, detects and discriminates dipole stimuli with the late...
19888887 - Illumination with monochromatic long-wavelength light promotes myopic shift and ocular ...
Publication Detail:
Type:  Journal Article     Date:  2010-10-15
Journal Detail:
Title:  Journal of anesthesia     Volume:  24     ISSN:  1438-8359     ISO Abbreviation:  J Anesth     Publication Date:  2010 Dec 
Date Detail:
Created Date:  2010-12-14     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8905667     Medline TA:  J Anesth     Country:  Japan    
Other Details:
Languages:  eng     Pagination:  888-92     Citation Subset:  IM    
Affiliation:
Intensive Care Unit, Osaka University Hospital, 2-15, Yamadaoka, Suita, Osaka, 565-0871, Japan, toooldwaterboys@yahoo.co.jp.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Estimation of the tourism climate in the Hunter Region, Australia, in the early twenty-first century...
Next Document:  Severe burn injury associated with misuse of forced-air warming device.