Document Detail


Evaluation of a new transcardiac conductance method for continuous on-line measurement of left ventricular volume.
MedLine Citation:
PMID:  10834719     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: To evaluate a new, less invasive, conductance method to measure continuous on-line left ventricular volume. End-systolic and end-diastolic volumes obtained with this transcardiac conductance method were compared with simultaneous measurements using the conventional intracardiac conductance catheter. DESIGN: Controlled animal study. SETTING: Research laboratory in a university hospital. SUBJECTS: Six sheep. INTERVENTIONS: Anesthetized sheep were instrumented and inotropic condition was varied by beta-receptor stimulation (5 microg/kg/min of dobutamine) and beta-receptor blockade (1 mg/kg of propranolol). In each condition (control, dobutamine, repeat control, propranolol), ventricular volume was varied over a wide range by gradual preload reduction using a vena caval balloon catheter. MEASUREMENTS AND MAIN RESULTS: We compared the two methods by performing linear regression analysis on simultaneous end-systolic and end-diastolic volumes obtained during gradual caval occlusions. We statistically analyzed the intercepts, slopes, and correlation coefficients of the regression equations relating the transcardiac and conductance catheter measurements to determine the effects of interanimal variability, inotropic condition, and cardiac phase on the relationship between the two methods. The results show an excellent linear correlation between the two methods (mean intercept, -1.82+/-1.24 mL; mean slope, 0.787+/-0.024 and r2 = .94). Both slope and intercept of the relationship between the two methods show a significant interanimal and cardiac phase related variability but no significant dependence on inotropic condition. CONCLUSIONS: The significant interanimal variability indicates that the new method requires individual calibration in each subject. However, the small variability of the regression coefficients with changes in condition indicates that after initial calibration, end-systolic and end-diastolic volume can be followed accurately even in the presence of large changes in volume and inotropic state. This new method may facilitate quantitative continuous assessment of cardiac function in clinical practice, for example, in the intensive care unit.
Authors:
P Steendijk; J W Lardenoye; E T van der Velde; M J Schalij; J Baan
Related Documents :
3249269 - Effect of heart rate on the end-systolic wall stress-mean velocity of circumferential f...
10423349 - Concentration and elongation of attached cross-bridges as pressure determinants in a ve...
476739 - Afterload as a predeterminant of haemodynamics and segmental wall motion following coro...
3974129 - Roles of coronary circulation as a determinant of the left ventriculo-arterial interact...
3578059 - Cardiac response to daily activities and exercise in normal subjects assessed by an amb...
7571389 - The effects of medetomidine on cardiac contractility in autonomically blocked dogs.
8205519 - The center of pressure index in the evaluation of foot orthoses in shoes.
14561709 - Uroguanylin knockout mice have increased blood pressure and impaired natriuretic respon...
22577769 - Optic disc fluorescein leakage and intraocular pressure in primary open-angle glaucoma.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Critical care medicine     Volume:  28     ISSN:  0090-3493     ISO Abbreviation:  Crit. Care Med.     Publication Date:  2000 May 
Date Detail:
Created Date:  2000-06-08     Completed Date:  2000-06-08     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  0355501     Medline TA:  Crit Care Med     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  1599-606     Citation Subset:  AIM; IM    
Affiliation:
Leiden University Medical Centre, Department of Cardiology, The Netherlands.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Calibration
Diagnosis, Computer-Assisted / instrumentation*
Diastole / physiology
Electrocardiography / instrumentation*
Electrodes
Heart Catheterization / instrumentation*
Myocardial Contraction / physiology
Online Systems / instrumentation*
Sensitivity and Specificity
Sheep
Stroke Volume / physiology*
Systole / physiology
Ventricular Function, Left / physiology*
Comments/Corrections
Comment In:
Crit Care Med. 2000 May;28(5):1686-7   [PMID:  10834754 ]