Evaluation in volunteers of the VIA V-ABG automated bedside blood gas, chemistry, and hematocrit monitor.
MedLine Citation:
PMID:  9951759     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: To evaluate the VIA V-ABG (VIA Medical Corp.) point-of-care blood gas and chemistry monitor in healthy human volunteers, with particular emphasis on the measurement of blood gases. METHODS: Experimental conditions were varied by intermittently subjecting volunteers to either isocapnic hypercapnia (end-tidal (ET), PETCO2 = 50+/-2 mmHg, ETPO2 = 130+/-5 mmHg) or isocapnic hypoxia (PETCO2 = 42+/-2, PETO2 + 45+/-2 mmHg) in addition to room air breathing. Measurements by the VIA V-ABG device were compared with paired samples and measurements performed by two ABL Radiometers (505 and 500). Analysis of results includes bias and precision plots and comparison of results with minimal performance criteria as established by CLIA. RESULTS: Nineteen volunteers yielded 222 matched samples. The range of values were 7.32-7.61 for pH, 20.9-51.6 mmHg for PCO2, 27.9-184.5 mmHg for PO2, 134-141 mmol/l for Na, 3.1-4.1 mmol/l for K, and 30.0-50.4% for hematocrit. Bias and precision (+/-2 sd) for pH was 0.01 and 0.04, for PCO2 was 0.4 and 4.8, for PO2 was 1.0 and 17.0, for Na was -0.3 and 5.2, for K was 0.1 and 0.2, and for Hct was 2.0 and 5.4. CONCLUSIONS: Over the range of blood gas values assessed, blood gas measurements by the VIA V-ABG device were clinically acceptable and met minimal performance criteria utilizing current Medicare CLIA proficiency standards. Performance criteria were also met by the VIA V-ABG device for Na, K, and Hct measurements but the range of values was too narrow to allow characterization of clinical acceptability. The VIA V-ABG device appears to perform well compared with the results which have been published for other point-of-care devices. Comparison between different studies investigating point-of-care devices is difficult due to several factors (range of values measured, comparison device, population studied, etc.). Some of these instruments, including the VIA V-ABG device, may serve quite well as point-of-care devices to perform certain tests at the bedside. Whether or not any of these devices can substitute for traditional laboratory blood gas and chemistry measurements remains an issue that is not adequately studied.
Authors:
P L Bailey; S W McJames; M L Cluff; D T Wells; J A Orr; D R Westenskow; S E Kern
Related Documents :
4015939 - Determination of the partial pressure of halothane (or isoflurane) in blood.
23369379 - Study of mo (vi) removal from aqueous solution: application of different mathematical m...
7978199 - Vaporization of isoflurane by liquid infusion.
1737999 - Dissimilar aluminum and gallium permeation of the blood-brain barrier demonstrated by i...
9951759 - Evaluation in volunteers of the via v-abg automated bedside blood gas, chemistry, and h...
15578469 - Accuracy and performance of a modified continuous intravascular blood gas monitoring de...
15542729 - Neuromechanism of developing methamphetamine psychosis: a neuroimaging study.
8306459 - Simple mixing technique for the preparation of blood samples of known oxygen saturation.
10379499 - Postnatal change of renal artery blood flow velocity and its relationship with urine vo...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Journal of clinical monitoring and computing     Volume:  14     ISSN:  1387-1307     ISO Abbreviation:  J Clin Monit Comput     Publication Date:  1998 Jul 
Date Detail:
Created Date:  1999-04-22     Completed Date:  1999-04-22     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9806357     Medline TA:  J Clin Monit Comput     Country:  NETHERLANDS    
Other Details:
Languages:  eng     Pagination:  339-46     Citation Subset:  IM    
Affiliation:
Department of Anesthesiology, University of Utah Health Sciences Center, Salt Lake City 84132, USA. pbailey@anesth.med.utah.edu
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Blood Chemical Analysis / instrumentation*
Blood Gas Analysis / instrumentation*
Female
Hematocrit / instrumentation*
Humans
Male
Monitoring, Physiologic / instrumentation*
Point-of-Care Systems

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Hydraulic analog for simultaneous representation of pharmacokinetics and pharmacodynamics: applicati...
Next Document:  Sampling intervals to record severe hypotensive and hypoxic episodes in anesthetised patients.