Document Detail


Evaluation of blood access dysfunction based on a wavelet transform analysis of shunt murmurs.
MedLine Citation:
PMID:  16807812     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
We investigated shunt murmurs based on wavelet transform analysis as a new method for assessing vascular access function. In the present study, in patients with venous stenosis near an arteriovenous fistula (A-V fistula), a sensor was placed at different positions around the stenosis and shunt murmur signals obtained using a measurement system were subjected to time-frequency analysis based on wavelet transforms. The shunt murmurs obtained from the stenotic region closely represented some features of murmurs that are often referred to as "high-pitch" murmurs in the clinical setting. In contrast, shunt murmurs obtained about 5 cm downstream of the stenotic region closely represented some features of murmurs that are often referred to as "low-pitch" murmurs in the clinical setting. Furthermore, with the aim of extending the lifespan of arteriovenous grafts (A-V grafts) by detecting and treating stenotic lesions before the A-V graft becomes occluded, we evaluated the possibility of utilizing the present shunt murmur analysis for monitoring stenosis in such A-V grafts. When shunt murmurs from patients with A-V grafts were analyzed, the results suggested that the blood flow through the venous anastomosis of the graft was the most turbulent. This present method whereby blood flow in an A-V fistula is assessed based on the frequency distribution on a time-frequency plane by wavelet transform analysis is advantageous because findings are not markedly affected by sensor attachment. Furthermore, because the sensor is attached using an adhesive collar, measurements can be taken over a short period of time before each dialysis session.
Authors:
Toshio Sato; Kiichi Tsuji; Norimichi Kawashima; Tetsuzo Agishi; Hiroshi Toma
Publication Detail:
Type:  Evaluation Studies; Journal Article    
Journal Detail:
Title:  Journal of artificial organs : the official journal of the Japanese Society for Artificial Organs     Volume:  9     ISSN:  1434-7229     ISO Abbreviation:  J Artif Organs     Publication Date:  2006  
Date Detail:
Created Date:  2006-06-29     Completed Date:  2006-11-21     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9815648     Medline TA:  J Artif Organs     Country:  Japan    
Other Details:
Languages:  eng     Pagination:  97-104     Citation Subset:  IM    
Affiliation:
Department of Clinical Engineering, Faculty of Biomedical Engineering, Toin University of Yokohama, Yokohama 225-8502, Japan. toshio@cc.toin.ac.jp
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Anastomosis, Surgical*
Blood*
Heart Murmurs*
Humans

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Nanopillar sheets as a new type of cell culture dish: detailed study of HeLa cells cultured on nanop...
Next Document:  Development of a highly responsive needle-type glucose sensor using polyimide for a wearable artific...