| Evaluation of the VITROS 5,1 FS chemistry system in a large pediatric laboratory. | |
| | |
MedLine Citation:
|
PMID: 17323821 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
The VITROS 5,1 FS analytical system was evaluated. The measurement of the serum proteins and HDL cholesterol with the newly developed MicroTip technology was compared with other nephelometric and turbidimetric methods. The within-run imprecision for apolipoprotein A1, apolipoprotein B, complement 3, complement 4, transferrin, immunoglobulin A, immunoglobulin M, immunoglobulin G and HDL cholesterol showed coefficients of variation between 0.4% and 4.7%. The day-to-day imprecision showed coefficients of variation between 0.5% and 4.0%. The correlation with the comparative methods (nephelometric or turbidimetric) was good, with Pearson correlation coefficients of between 0.959 and 0.995. The comparison of apolipoprotein Al showed a slope of about 0.76. An explanation for the difference could not be found. The VITROS 5,1 FS analytical system is a reliable routine instrument which fits into a large pediatric laboratory of a university. |
| | |
Authors:
|
Kurt Herkner; Rosmarie Reiter-Reisacher; Oswald Sonntag |
Publication Detail:
|
Type: Journal Article |
Journal Detail:
|
Title: Clinical laboratory Volume: 53 ISSN: 1433-6510 ISO Abbreviation: Clin. Lab. Publication Date: 2007 |
Date Detail:
|
Created Date: 2007-02-27 Completed Date: 2007-03-15 Revised Date: 2007-11-15 |
Medline Journal Info:
|
Nlm Unique ID: 9705611 Medline TA: Clin Lab Country: Germany |
Other Details:
|
Languages: eng Pagination: 17-26 Citation Subset: IM |
Affiliation:
|
Allgemeines Krankenhaus, Universitäts-Kinderklinik, Vienna, Austria. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Apolipoproteins
/
blood Blood Chemical Analysis / instrumentation*, methods* Cholesterol, HDL / blood Complement System Proteins / analysis Evaluation Studies as Topic Humans Immunoglobulins / blood Nephelometry and Turbidimetry / instrumentation*, methods* Pediatrics / instrumentation, methods Regression Analysis Transferrin / analysis |
| Chemical | |
Reg. No./Substance:
|
0/Apolipoproteins; 0/Cholesterol, HDL; 0/Immunoglobulins; 11096-37-0/Transferrin; 9007-36-7/Complement System Proteins |