Document Detail


Evaluation of RAPID(®) 5 Access software for examination of capsule endoscopies and reading of the capsule by an endoscopy nurse.
MedLine Citation:
PMID:  20821237     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
BACKGROUND: Since capsule endoscopy (CE) is time consuming, one possible cost-effective strategy could be the use of an expert endoscopic assistant and available software to select images. Aims were to examine the clinical utility of RAPID(®) 5 Access software and find the optimum setting mode for reading. We also evaluated whether a nurse could preview the CE video and detect significant lesions accurately.
METHODS: The capsule images in 14 volunteers with known mucosal injury induced by low dose aspirin and in 30 patients who were known to have a variety of significant lesions were selected. Using three setting modes of RAPID(®) 5 Access software, the detection rate and reading time for CE images by two well-trained physicians and one expert nurse were compared.
RESULTS: There was no significant difference in detection rate among the three readers. The detection rate using Quickview RAPID(®) 5 Access was significantly higher than that using RAPID(®) Reader version 4.1. Comparison among the three modes of RAPID(®) 5 Access showed that auto mode as well as displaying a single image at 12 fps was superior in the detection rate of denuded redness, while its reading time was longer compared to the other modes. Some significant lesions were not detected by using Quickview and Quadview modes.
CONCLUSIONS: RAPID(®) 5 Access improves diagnostic yield, reducing reading time; however, it is still unacceptable because of the diagnostic miss rate and may be useful as an ancillary reading tool. Developing further improved software and training expert assistants for reading capsule images are necessary.
Authors:
Akiko Shiotani; Keisuke Honda; Makiko Kawakami; Takahisa Murao; Hiroshi Matsumoto; Ken-Ichi Tarumi; Hiroaki Kusunoki; Jiro Hata; Ken Haruma
Related Documents :
22394787 - 65 three-dimensional chemical imaging of a whole pig fetus by desorption electrospray i...
10278217 - Do you understand the role of software? it's the key to performance of laboratory compu...
11573887 - A portable system for the measurement of the temporal parameters of gait.
22502577 - Use of ambient light in remote photoplethysmographic systems: comparison between a high...
10125547 - Improving customer relations and telephone response in abbott northwestern hospital's l...
23930817 - Analytical cone-beam spect reconstruction algorithm with non-uniform attenuation for ge...
Publication Detail:
Type:  Journal Article     Date:  2010-09-07
Journal Detail:
Title:  Journal of gastroenterology     Volume:  46     ISSN:  1435-5922     ISO Abbreviation:  J. Gastroenterol.     Publication Date:  2011 Feb 
Date Detail:
Created Date:  2011-02-18     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9430794     Medline TA:  J Gastroenterol     Country:  Japan    
Other Details:
Languages:  eng     Pagination:  138-42     Citation Subset:  IM    
Affiliation:
Department of Internal Medicine, Kawasaki Medical School, 577 Matsushima, Kurashiki, Okayama, 701-0192, Japan, shiotani@med.kawasaki-m.ac.jp.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Clinicopathological features of severe and fulminant forms of autoimmune hepatitis.
Next Document:  Evaluation of efficacy and safety of fentanyl transdermal patch (Durogesic D-TRANS) in chronic pain.