Document Detail


Evaluation of high-strength orthopedic sutures: a head-to-head comparison.
MedLine Citation:
PMID:  20839707     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
The goal of this study was to determine whether a difference in cycles to failure or mode of failure would be observed among specimens of 3 high-strength suture materials, and whether different suture configurations would affect knot security. Ten representative specimens of Ethibond (Ethicon, Inc, Somerville, New Jersey), FiberWire (Arthrex, Inc, Naples, Florida), MaxBraid (Biomet, Inc, Warsaw, Indiana), and Orthocord (DePuy Orthopaedics, Warsaw, Indiana) were tied in 6 different knot configurations commonly used in orthopedic procedures. Each specimen was cyclically loaded between 9 and 180 N at a rate of 1 Hz until the specimen failed or reached a maximum of 3500 cycles. Each suture material was subjected to tensile loading until failure at a rate of 1.25 mm/s. The 3 most secure knots all included the 3 reverse half-hitch on alternating posts (3-RHAP) configuration. All specimens tied with these 3 knot types failed by suture rupture. All knots using the overhand with 3 of the same half-hitches on the same post (O-3SHSP) configuration failed by knot slippage regardless of suture material. When the 3 strongest knots were combined, FiberWire resisted a significantly greater number of fatigue cycles than Orthocord or MaxBraid. In the single load to failure tests, Orthocord, FiberWire, and MaxBraid all had significantly higher ultimate strength than Ethibond. Knots using the 3-RHAP configuration provide security superior to that of those without this configuration. All 3 high-strength sutures tested outperformed Ethibond in single load to failure testing, with FiberWire resisting the greatest number of cycles. Postoperative strength and reliability of a soft tissue repair is inherently dependent on the properties of the suture materials used.
Authors:
Timothy Miller; Jeff Feinblatt; John Craw; Alan Litsky; David Flanigan
Related Documents :
25281647 - The effect of osteotomy dimension on osseointegration to resorbable media-treated impla...
21288107 - Profound deafness in childhood.
415277 - Cheek mucosa response to silk, cotton, and nylon suture materials.
3551877 - The sustained release of cefotaxim from a fibrin-cefotaxim compound in treatment of ost...
727907 - The shear strength of trabecular bone from the femur, and some factors affecting the sh...
9108127 - Changes in anandamide levels in mouse uterus are associated with uterine receptivity fo...
Publication Detail:
Type:  Journal Article     Date:  2010-09-07
Journal Detail:
Title:  Orthopedics     Volume:  33     ISSN:  1938-2367     ISO Abbreviation:  Orthopedics     Publication Date:  2010 Sep 
Date Detail:
Created Date:  2010-09-15     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7806107     Medline TA:  Orthopedics     Country:  United States    
Other Details:
Languages:  eng     Pagination:  674     Citation Subset:  IM    
Copyright Information:
Copyright 2010, SLACK Incorporated.
Affiliation:
Department of Orthopedics, The Ohio State University, Columbus, Ohio, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Effect of preoperative exercise on postoperative mobility in obese total joint replacement patients.
Next Document:  Effectiveness of external rotation immobilization in highly active young men with traumatic primary ...