Evaluation of EIT system performance.
MedLine Citation:
PMID:  21646713     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
An electrical impedance tomography (EIT) system images internal conductivity from surface electrical stimulation and measurement. Such systems necessarily comprise multiple design choices from cables and hardware design to calibration and image reconstruction. In order to compare EIT systems and study the consequences of changes in system performance, this paper describes a systematic approach to evaluate the performance of the EIT systems. The system to be tested is connected to a saline phantom in which calibrated contrasting test objects are systematically positioned using a position controller. A set of evaluation parameters are proposed which characterize (i) data and image noise, (ii) data accuracy, (iii) detectability of single contrasts and distinguishability of multiple contrasts, and (iv) accuracy of reconstructed image (amplitude, resolution, position and ringing). Using this approach, we evaluate three different EIT systems and illustrate the use of these tools to evaluate and compare performance. In order to facilitate the use of this approach, all details of the phantom, test objects and position controller design are made publicly available including the source code of the evaluation and reporting software.
Authors:
Mamatjan Yasin; Stephan Böhm; Pascal O Gaggero; Andy Adler
Related Documents :
21849403 - Using the nema nu 4 pet image quality phantom in multipinhole small-animal spect.
21319173 - Cellular analysis by open-source software for affordable cytometry.
11876323 - Theory of near-field magneto-optical imaging.
Publication Detail:
Type:  Journal Article     Date:  2011-06-07
Journal Detail:
Title:  Physiological measurement     Volume:  32     ISSN:  1361-6579     ISO Abbreviation:  Physiol Meas     Publication Date:  2011 Jul 
Date Detail:
Created Date:  2011-06-07     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9306921     Medline TA:  Physiol Meas     Country:  England    
Other Details:
Languages:  eng     Pagination:  851-65     Citation Subset:  IM    
Affiliation:
Systems and Computer Engineering, Carleton University, Ottawa, Canada.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Minimizing EIT image artefacts from mesh variability in finite element models.
Next Document:  Measurement of regional pulmonary oxygen uptake-a novel approach using electrical impedance tomograp...