Document Detail


Estimation of normal hydration in dialysis patients using whole body and calf bioimpedance analysis.
MedLine Citation:
PMID:  21646705     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
Prescription of an appropriate dialysis target weight (dry weight) requires accurate evaluation of the degree of hydration. The aim of this study was to investigate whether a state of normal hydration (DW(cBIS)) as defined by calf bioimpedance spectroscopy (cBIS) and conventional whole body bioimpedance spectroscopy (wBIS) could be characterized in hemodialysis (HD) patients and normal subjects (NS). wBIS and cBIS were performed in 62 NS (33 m/29 f) and 30 HD patients (16 m/14 f) pre- and post-dialysis treatments to measure extracellular resistance and fluid volume (ECV) by the whole body and calf bioimpedance methods. Normalized calf resistivity (ρ(N)(,5)) was defined as resistivity at 5 kHz divided by the body mass index. The ratio of wECV to total body water (wECV/TBW) was calculated. Measurements were made at baseline (BL) and at DW(cBIS) following the progressive reduction of post-HD weight over successive dialysis treatments until the curve of calf extracellular resistance is flattened (stabilization) and the ρ(N)(,5) was in the range of NS. Blood pressures were measured pre- and post-HD treatment. ρ(N)(,5) in males and females differed significantly in NS. In patients, ρ(N)(,5) notably increased with progressive decrease in body weight, and systolic blood pressure significantly decreased pre- and post-HD between BL and DW(cBIS) respectively. Although wECV/TBW decreased between BL and DW(cBIS), the percentage of change in wECV/TBW was significantly less than that in ρ(N)(,5) (-5.21 ± 3.2% versus 28 ± 27%, p < 0.001). This establishes the use of ρ(N)(,5) as a new comparator allowing a clinician to incrementally monitor removal of extracellular fluid from patients over the course of dialysis treatments. The conventional whole body technique using wECV/TBW was less sensitive than the use of ρ(N)(,5) to measure differences in body hydration between BL and DW(cBIS).
Authors:
Fansan Zhu; Peter Kotanko; Garry J Handelman; Jochen G Raimann; Li Liu; Mary Carter; Martin K Kuhlmann; Eric Seibert; Edward F Leonard; Nathan W Levin
Related Documents :
11767315 - The influence of ringer's lactate and gelatin infusion on the internal fluid balance of...
22835615 - Use of continuous positive airway pressure reduces airway reactivity in adults with ast...
19328085 - Superwet anesthesia redefines large-volume liposuction.
21845435 - Squatting test: a dynamic postural manoeuvre to study baroreflex sensitivity.
18269185 - Biventricular function at high altitude: implications for regulation of stroke volume i...
25319815 - Dynamic clinical measurements of voluntary vaginal contractions and autonomic vaginal r...
Publication Detail:
Type:  Journal Article     Date:  2011-06-07
Journal Detail:
Title:  Physiological measurement     Volume:  32     ISSN:  1361-6579     ISO Abbreviation:  Physiol Meas     Publication Date:  2011 Jul 
Date Detail:
Created Date:  2011-06-07     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9306921     Medline TA:  Physiol Meas     Country:  England    
Other Details:
Languages:  eng     Pagination:  887-902     Citation Subset:  IM    
Affiliation:
Renal Research Institute, Yorkville Dialysis Center, 1555 3rd Avenue 218, New York, NY 10128, USA. Division of Nephrology and Hypertension, Beth Israel Medical Center, New York, NY, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Impedance sensor technology for cell-based assays in the framework of a high-content screening syste...
Next Document:  A fully parallel multi-frequency EIT system with flexible electrode configuration: KHU Mark2.