Errors in flow and pressure related to the arterial filter purge line.
MedLine Citation:
PMID:  10182117     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The purge line is a necessary component on arterial filters, although its presence may affect the amount of flow reaching the patient as well as the pump outlet pressure in the extracorporeal circuit. In-vitro and clinical studies conducted to investigate these effects with a commonly used purge line showed that at flows less than 1.5 L/min, rates for pediatric or infant patients, the purge line diverts as much as 40% of the intended pump flow away from the patient. A small diameter resistance tube connected in series with the purge line reduced purge flow such that over 80% of the pump flow reached the patient. Pressure monitored at the arterial filter port with the purge line open could be as much as 45 mmHg lower than the pressure measured with the purge line closed to the filter. Studies should be done to determine if the arterial filter purge line compromises flow to the patient, and if an additional resistance to the purge line is appropriate to reduce the flow through it.
Authors:
K Lee-Sensiba; N Azzaretto; C Carolina; N DiCarmine; D Hymowitz; S Kay; K Kooker; M Salogub; E Wong; Y Tamari
Related Documents :
19294307 - A fritless, eof microchip pump for high pressure pumping of aqueous and organic solvents.
12889397 - Intermittent pneumatic compression pump settings for the optimum reduction of oedema.
8694697 - Modification of a pivot bearing system on a compact centrifugal pump.
2308307 - Experimental analysis of pulsatile flow through elastic collapsible tubes: application ...
3462397 - Unilateral nasal resistance and asymmetrical body pressure.
25060557 - Central aortic pressure and pulsatility index in acute ischemic stroke.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  The Journal of extra-corporeal technology     Volume:  30     ISSN:  0022-1058     ISO Abbreviation:  J Extra Corpor Technol     Publication Date:  1998 Jun 
Date Detail:
Created Date:  1998-09-14     Completed Date:  1998-09-14     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  0267637     Medline TA:  J Extra Corpor Technol     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  77-82     Citation Subset:  T    
Affiliation:
Dept. of Surgery, North Shore University Hospital, New York University School of Medicine, Manhasset 11030, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Arteries
Blood Circulation / physiology
Blood Pressure / physiology
Child
Equipment Design
Extracorporeal Circulation / instrumentation*
Filtration / instrumentation*
Hemorheology
Humans
Infant
Intubation / instrumentation

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Do different heparin brands influence activated clotting times?
Next Document:  Hematological abnormalities in neonatal patients treated with extracorporeal membrane oxygenation (E...