Abstract/OtherAbstract:
|
BACKGROUND: The variability introduced by inconsistent placement of pressure transducers for invasive monitoring may result in significant measurement error. Our goals in this study were to quantify the degree of variation among health care providers and to identify a simple tool for reducing this error. METHODS: A sample of 50 perioperative health care providers was recruited and asked to place a transducer at the appropriate level for central venous pressure (CVP) monitoring on two separate occasions: first without any additional standardization tools and second with a laser level to guide transducer placement. The variability among providers was calculated, and the results between sessions compared. RESULTS: There was significant variation in transducer placement during both sessions, in some instances, of greater magnitude than a normal CVP value. The laser level did not significantly reduce this variation. CONCLUSION: There is significant variation in transducer placement among health care providers. This variation is not reduced by a laser level and must be considered when interpreting CVP data. Hospital- or institution-wide standardization of a zero-level should be considered. |
| | |
Authors:
|
Katie K Figg; Edward C Nemergut |
Related Documents
:
|
9503108 - Laboratory surveillance method for nosocomial clostridium difficile diarrhea. 10037488 - Characterization of ly231617 protection against hydrogen peroxide toxicity. 12595278 - Co2 transduction in avian intrapulmonary chemoreceptors is critically dependent on tran... 1492568 - The mediation of increased duodenal alkaline secretion in response to 10 mm hcl in the ... 16600418 - Disease management and health care reforms in germany - does more competition lead to l... 9826248 - Repetitively violent patients in psychiatric units. |
Publication Detail:
|
Type: Journal Article; Research Support, Non-U.S. Gov't |
Journal Detail:
|
Title: Anesthesia and analgesia Volume: 108 ISSN: 1526-7598 ISO Abbreviation: Anesth. Analg. Publication Date: 2009 Apr |
Date Detail:
|
Created Date: 2009-03-20 Completed Date: 2009-04-08 Revised Date: - |
Medline Journal Info:
|
Nlm Unique ID: 1310650 Medline TA: Anesth Analg Country: United States |
Other Details:
|
Languages: eng Pagination: 1209-11 Citation Subset: AIM; IM |
Affiliation:
|
Department of Anesthesiology, University of Virginia Health System, Charlottesville, VA 22908, USA. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Blood Pressure Determination*
/
instrumentation,
standards Blood Pressure Monitors Central Venous Pressure* Head-Down Tilt Humans Lasers Observer Variation Reproducibility of Results Supine Position Transducers, Pressure |