Document Detail


Error in central venous pressure measurement.
MedLine Citation:
PMID:  19299788     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: The variability introduced by inconsistent placement of pressure transducers for invasive monitoring may result in significant measurement error. Our goals in this study were to quantify the degree of variation among health care providers and to identify a simple tool for reducing this error. METHODS: A sample of 50 perioperative health care providers was recruited and asked to place a transducer at the appropriate level for central venous pressure (CVP) monitoring on two separate occasions: first without any additional standardization tools and second with a laser level to guide transducer placement. The variability among providers was calculated, and the results between sessions compared. RESULTS: There was significant variation in transducer placement during both sessions, in some instances, of greater magnitude than a normal CVP value. The laser level did not significantly reduce this variation. CONCLUSION: There is significant variation in transducer placement among health care providers. This variation is not reduced by a laser level and must be considered when interpreting CVP data. Hospital- or institution-wide standardization of a zero-level should be considered.
Authors:
Katie K Figg; Edward C Nemergut
Related Documents :
9503108 - Laboratory surveillance method for nosocomial clostridium difficile diarrhea.
10037488 - Characterization of ly231617 protection against hydrogen peroxide toxicity.
12595278 - Co2 transduction in avian intrapulmonary chemoreceptors is critically dependent on tran...
1492568 - The mediation of increased duodenal alkaline secretion in response to 10 mm hcl in the ...
16600418 - Disease management and health care reforms in germany - does more competition lead to l...
9826248 - Repetitively violent patients in psychiatric units.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Anesthesia and analgesia     Volume:  108     ISSN:  1526-7598     ISO Abbreviation:  Anesth. Analg.     Publication Date:  2009 Apr 
Date Detail:
Created Date:  2009-03-20     Completed Date:  2009-04-08     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  1310650     Medline TA:  Anesth Analg     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1209-11     Citation Subset:  AIM; IM    
Affiliation:
Department of Anesthesiology, University of Virginia Health System, Charlottesville, VA 22908, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Blood Pressure Determination* / instrumentation,  standards
Blood Pressure Monitors
Central Venous Pressure*
Head-Down Tilt
Humans
Lasers
Observer Variation
Reproducibility of Results
Supine Position
Transducers, Pressure

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  The impact of an electronic reminder on the use of alarms after separation from cardiopulmonary bypa...
Next Document:  Telemedicine consultation and monitoring for pediatric liver transplant.