The TaperGuard™ Endotracheal Tube Intracuff Pressure Increase Is Less Than That of the Hi-Lo™ Tube During Nitrous Oxide Exposure: A Model Trachea Study.
MedLine Citation:
PMID:  23400983     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
BACKGROUND:Studies have compared sealing effects of the newly developed tapered endotracheal tube cuff with the conventional cylindrical cuff. In this study, we compared the difference between cuffs with regard to the increase in intracuff pressure during nitrous oxide (N(2)O) exposure.METHODS:Two types of cuffs were studied using a model trachea connected to a mechanical test lung: high-volume, low-pressure cuff (Mallinckrodt Hi-Lo(TM), Covidien, Dublin, Ireland) and tapered cuff (Mallinckrodt TaperGuardTM, Covidien). The intracuff pressure was set at 20 cm H(2)O, and the increase in pressure was measured during mechanical ventilation using 66% N(2)O. Intracuff pressures were recorded after 5, 10, 15, 30, 45, and 60 minutes of exposure to N(2)O.RESULTS:The intracuff pressure recorded during the first 15 minutes of N(2)O exposure in high-volume, low-pressure cuffs was significantly higher than tapered cuffs (2-way repeated-measures analysis of variance, P < 0.0001 for internal diameters [IDs] 7.0 and 7.5 mm, P = 0.0004 for ID 8.0 mm, P = 0.0013 for ID 8.5 mm), and there were also statistically significant differences regarding interaction of time and cuff type (P < 0.0001 for IDs 7.0, 7.5, 8.0, and 8.5 mm). The difference in mean cuff pressures among groups after 10 minutes of N(2)O exposure was -18.5 (SE, 1.4; 99% confidence interval, -22.8 to -14.2; P < 0.0001) for ID 7.5 mm. Tapered endotracheal tube cuffs sealed the trachea with fewer dimples on the carina side of the cuff. Dimples on the cuff surface probably increase the surface for N(2)O diffusion. Therefore, fewer dimples result in a smaller surface area through which N(2)O can diffuse.CONCLUSION:During general anesthesia with N(2)O, the intracuff pressure of tapered endotracheal tube cuffs did not increase as rapidly as it did in conventional high-volume, low-pressure cuffs. The pressure in both types of cuffs increased rapidly when exposed to 66% N(2)O, and hence continuous or frequent monitoring is recommended.
Authors:
Sayaka Tsuboi; Tetsuya Miyashita; Yoshikazu Yamaguchi; Yoshiko Yamamoto; Kentaro Sakamaki; Takahisa Goto
Related Documents :
12960063 - Administration of testosterone is associated with a reduced susceptibility to myocardia...
3591243 - The arrangement of endoplasmic reticulum and proteosynthesis in spinal ganglia neurons ...
17209693 - Cardiac autonomic reactivity and recovery in predicting carotid atherosclerosis: the ca...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2013-2-11
Journal Detail:
Title:  Anesthesia and analgesia     Volume:  -     ISSN:  1526-7598     ISO Abbreviation:  Anesth. Analg.     Publication Date:  2013 Feb 
Date Detail:
Created Date:  2013-2-12     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  1310650     Medline TA:  Anesth Analg     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
From the *Department of Anesthesiology, Yokohama City University Hospital, Yokohama; and †Department of Biostatistics and Epidemiology, Graduate School of Medicine, Yokohama City University, Yokohama, Japan.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  The Impact of Bispectral Index Versus End-Tidal Anesthetic Concentration-Guided Anesthesia on Time t...
Next Document:  A feasibility study using the VivaSight Single Lumen™ to intubate the trachea through the Fastrach l...