Document Detail


Endotoxin levels in steam and reservoirs of table-top steam sterilizers.
MedLine Citation:
PMID:  11828908     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
PURPOSE: To document endotoxin levels in "Statim" cassette sterilizer reservoirs and in steam delivered to the cassette in the unwrapped instrument cycle. To document endotoxin levels in sterilizer reservoir water using different management protocols. METHODS: Endotoxin levels were determined using the Limulus Amebocyte Lysate test. Endotoxin preparations were from Escherichia coli and Ralstonia pickettii. All samples were collected in depyrogenated glassware and stored at -20 degrees C until assayed. RESULTS: The majority of water samples contained < 1.0 Endotoxin Unit (EU)/ml. The highest level found in sterilizers in clinical use was 5.3 EU/ml. Endotoxin was not detected in steam condensate within the limits of the assay. When the endotoxin level in the reservoir water was experimentally enhanced to 200 EU/ml, cassette steam condensate endotoxin levels were from 0.5% to 5% of the reservoir level. Daily and weekly emptying of the cassette reservoir consistently yielded low endotoxin levels as did monthly emptying, but with the latter there was a trend toward higher levels that favors weekly emptying as a precautionary measure. CONCLUSIONS: Endotoxin levels in the reservoirs of 23 sterilizers involving 240 samplings were never high enough to yield detectable endotoxin levels in steam in the sterilizer cassette. Regular weekly emptying of sterilizer reservoirs would eliminate the risk of endotoxin transfer during steam sterilization.
Authors:
James L Whitby; Victoria M Hitchins
Related Documents :
847568 - A laboratory investigation into bilirubin binding in neonates.
3929608 - Nonpredictive value of measurements of delta optical density at 450 nm in ss disease.
7678928 - 'idiopathic' jaundice in sardinian full-term newborn infants: a multivariate study.
11431858 - Comparison of conventional phototherapy and fiberoptic phototherapy in the very-low-bod...
10845438 - Effect of early ambroxol treatment on lung functions in mechanically ventilated preterm...
2589038 - International collaborative effort (ice) on birth weight, plurality, perinatal, and inf...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Journal of refractive surgery (Thorofare, N.J. : 1995)     Volume:  18     ISSN:  1081-597X     ISO Abbreviation:  J Refract Surg     Publication Date:    2002 Jan-Feb
Date Detail:
Created Date:  2002-02-06     Completed Date:  2002-11-07     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9505927     Medline TA:  J Refract Surg     Country:  United States    
Other Details:
Languages:  eng     Pagination:  51-7     Citation Subset:  IM    
Affiliation:
Department of Microbiology and Immunology, University of Western Ontario, London, Canada. jwhitby@uwo.ca
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Ambulatory Care
Endotoxins / analysis*
Equipment Contamination*
Escherichia coli
Gram-Negative Aerobic Rods and Cocci
Limulus Test
Sterilization / instrumentation*
Water Microbiology*
Chemical
Reg. No./Substance:
0/Endotoxins

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Risk factors for insufficient fixation of microkeratome during laser in situ keratomileusis.