Document Detail


Emergent Reconnection of a Transected Left Ventricular Assist Device Driveline.
MedLine Citation:
PMID:  25220024     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
BACKGROUND: Recent clinical advances with new mechanical circulatory systems have led to additional strategies in the treatment of end-stage heart failure. The third-generation HeartWare Left Ventricular Device (LVAD) System utilizes a blood pump and a driveline (cable) that exits the patient's skin connecting the implanted pump to an externally worn controller. We report a rare case of a HeartWare LVAD driveline rewiring after accidental (presumed) transection of the driveline system.
CASE REPORT: A 67-year-old male with a medical history of ischemic cardiomyopathy status post HeartWare LVAD implantation presented to the emergency department (ED) after acute LVAD failure. On the morning of presentation, he attempted to cut the paper tape off of his adult diaper with scissors and accidentally (presumed) and unwitnessed severed the driveline system. The patient immediately went into cardiac arrest and was transported to a regional medical center. On arrival, he exhibited no appreciable vital signs and was subsequently intubated, vascular access was placed, and inotropic support initiated. The emergency physician individually stripped and reconnected the color-coded driveline wires using multiple hemostats, electrical tape, and cardboard, which resulted in regeneration of positive LVAD flows. WHY SHOULD AN EMERGENCY PHYSICIAN BE AWARE OF THIS?: VAD patients will present in extremis typically to the ED with manifestations of pump dysfunction ranging from diminished flows needing fluid management or pump adjustments to full pump failure manifesting as cardiogenic shock, needing rapid resuscitation and transfer to a cardiothoracic surgical unit with on-site VAD-perfusion specialists.
Authors:
Efrain Israel Cubillo; Ricardo A Weis; Harish Ramakrishna
Related Documents :
19665014 - Neuroprotective effects of testosterone upon cardiac sympathetic function in rats with ...
11542384 - Fractal fluctuations in cardiac time series.
19729694 - Heart rate recovery as a predictor of mortality with or without revascularization.
8749874 - Plasma norepinephrine, myocardial damage and left ventricular systolic function in chag...
1370924 - Pharmacokinetics of tissue-type plasminogen activator during acute myocardial infarctio...
19114204 - Effect of rosuvastatin pretreatment on myocardial damage after coronary surgery: a rand...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2014-9-11
Journal Detail:
Title:  The Journal of emergency medicine     Volume:  -     ISSN:  0736-4679     ISO Abbreviation:  J Emerg Med     Publication Date:  2014 Sep 
Date Detail:
Created Date:  2014-9-15     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8412174     Medline TA:  J Emerg Med     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Copyright Information:
Copyright © 2014 Elsevier Inc. All rights reserved.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Extracorporeal Life Support and Plasmapheresis in a Case of Severe Polyintoxication.
Next Document:  Association of Mandated Language Access Programming and Quality of Care Provided by Mental Health Ag...