Document Detail


Emergency reinforcement of cracked paracorporeal blood pumps.
MedLine Citation:
PMID:  15672799     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
In a patient with biventricular paracorporeal cardiac support, severe cracks in the outer pump housing occurred for unknown reasons (possibly because of the application of solvent). Pump replacement was considered but estimated to be a very high risk for this particular patient. Therefore, it was decided to glue the ventricles. Gluing was done after pilot tests in a two stage procedure: first, with a special glue that hardens under exposure to ultraviolet light and then with a two component low exothermic epoxy resin. The procedure resulted in very satisfying stability for the remainder of the bridging period, and the patient could be transplanted successfully 59 days after the event. It is concluded that a carefully selected gluing procedure (caveats: with previously tested glue, a validated procedure, and as a last resort only) may bring less risk in such rare cases of pump cracks than a pump exchange.
Authors:
Heinrich Schima; Christian Reindl; Martin Stoiber; Wilfried Röthy; Georg Wieselthaler
Related Documents :
9506219 - Delayed-type hypersensitivity to contact allergens in psoriasis. a clinical evaluation.
20152899 - Transdermal fentanyl matrix patches matrifen and durogesic dtrans are bioequivalent.
10344479 - Late-type allergy to the x-ray contrast medium solutrast (iopamidol).
20536479 - Variability in patch test reactions--first report from the norwegian patch test registry.
17239989 - Results of patch testing to a corticosteroid series: a retrospective review of 1188 pat...
16536809 - Late reactions to the patch-test preparations para-phenylenediamine and epoxy resin: a ...
11178399 - animal tests in pharmacopoeia monographs. current state and possibilities for improvements
15460659 - Bioinformatical results based on statistical methods for endectocide products against f...
17411349 - The integrated acute systemic toxicity project (acutetox) for the optimisation and vali...
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  ASAIO journal (American Society for Artificial Internal Organs : 1992)     Volume:  50     ISSN:  1058-2916     ISO Abbreviation:  ASAIO J.     Publication Date:    2004 Nov-Dec
Date Detail:
Created Date:  2005-01-27     Completed Date:  2005-06-30     Revised Date:  2006-05-23    
Medline Journal Info:
Nlm Unique ID:  9204109     Medline TA:  ASAIO J     Country:  United States    
Other Details:
Languages:  eng     Pagination:  621-3     Citation Subset:  IM    
Affiliation:
Department of Biomedical Engineering and Physics, Medical University of Vienna, Vienna, Austria.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adhesives / pharmacology
Adult
Emergency Medical Services*
Epoxy Resins / pharmacology
Equipment Failure
Female
Heart Diseases / surgery
Heart Transplantation
Heart-Assist Devices / adverse effects*
Humans
Time Factors
Chemical
Reg. No./Substance:
0/Adhesives; 0/Epoxy Resins

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Lymphocyte's activation and apoptosis after coronary artery bypass graft: a comparative study of two...
Next Document:  A single complementary-sense transcript of a geminiviral DNA beta satellite is determinant of pathog...