Electrospray ionization on porous spraying tips for direct sample analysis by mass spectrometry: enhanced detection sensitivity and selectivity using hydrophobic/hydrophilic materials as spraying tips.
MedLine Citation:
PMID:  23418151     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
RATIONALE: Despite various porous materials having been widely adopted as spraying tips for direct sample analysis using electrospray ionization mass spectrometry (ESI-MS), the effect of surface property and porosity of spraying tip materials on their analytical performances is not clear. Investigation of their relationships could provide insight into the proper choice and/or design of spraying tip materials for direct sample analysis.
METHODS: The effect of spraying tip materials with different polarities, including polyester and polyethylene (hydrophobic) and wood (hydrophilic), on the detection sensitivity for a variety of compounds, and on the ESI onset voltage, were studied using ESI-MS. The porosity of each type of spraying tip was characterized by scanning electron microscopy (SEM). Factors governing the detection sensitivity were determined based on the correlation of the detection sensitivity to the ESI onset voltage, the polarity, and the porosity of the spraying tip materials.
RESULTS: Hydrophobic tips (i.e., polyester and polyethylene) show better detection sensitivity for polar compounds but not for non-polar compounds, while hydrophilic tips (wooden tips) show the opposite effect. This phenomenon could be due to the difference in interaction between the analytes and the tips, causing the analytes to adsorb on the tip to different extents. In addition, the micro-porous nature of the tips could facilitate solvent diffusion for transporting analytes to the tip and maintain a stable spray for recording MS data. With the proper choice of spraying tip materials, trace amount of analytes at the picomole level can be detected with minimal sample pretreatment.
CONCLUSIONS: Both the polarity and the porosity of the spraying tip materials could significantly affect detection sensitivity for a wide variety of analytes. With proper choice of spraying tip material, ESI on a porous spraying tip could be a sensitive method for the direct analysis of daily life samples. Copyright © 2013 John Wiley & Sons, Ltd.
Authors:
Melody Yee-Man Wong; Ho-Wai Tang; Sin-Heng Man; Ching-Wan Lam; Chi-Ming Che; Kwan-Ming Ng
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Rapid communications in mass spectrometry : RCM     Volume:  27     ISSN:  1097-0231     ISO Abbreviation:  Rapid Commun. Mass Spectrom.     Publication Date:  2013 Mar 
Date Detail:
Created Date:  2013-02-18     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8802365     Medline TA:  Rapid Commun Mass Spectrom     Country:  England    
Other Details:
Languages:  eng     Pagination:  713-21     Citation Subset:  IM    
Copyright Information:
Copyright © 2013 John Wiley & Sons, Ltd.
Affiliation:
Department of Chemistry, Open Laboratory of Chemical Biology of the Institute of Molecular Technology for Drug Discovery and Synthesis, and State Key Laboratory of Synthetic Chemistry, The University of Hong Kong, Pokfulam Road, Hong Kong SAR, P.R. China.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Effects of organic matter on carbonate stable isotope ratios (?(13) C, ?(18) O values) - implication...
Next Document:  Liquid chromatography/electrospray ionization quadrupole time-of-flight mass spectrometry for the an...