Document Detail


Electrical safety during electrophysiological testing.
MedLine Citation:
PMID:  2456543     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
This report describes the inadvertent induction of non-sustained atrial and ventricular arrhythmias due to the malfunction of a programmable cardiac stimulator. The malfunction occurred when line power resumed after a brief municipal power outage ("blackout") during an invasive electrophysiological study. The stimulator spontaneously delivered very high frequency pulses through the electrode catheter to the myocardium which resulted in atrial and ventricular arrhythmias. During bench testing, the stimulator delivered a continuous train of high frequency output pulses (greater than or equal to 1 mA) when line voltage resumed normal level after it had dropped below 65 VAC. Electrical safety during electrophysiological testing requires a stimulator design which is immune to altered operating conditions, and which shuts down if abnormal operating or output conditions are detected.
Authors:
S A Rubin; P C Nalos; S Solingen; J Whiting; E Gang
Related Documents :
10379363 - Minoxidil-induced cardiac hypertrophy in guinea pigs.
23642963 - The effect of exercise repetition on the frequency characteristics of motor output forc...
6592523 - Temporomandibular joint function and morphology: observations on the spectra of normalcy.
Publication Detail:
Type:  Case Reports; Journal Article; Research Support, U.S. Gov't, P.H.S.    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  11     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  1988 May 
Date Detail:
Created Date:  1988-08-31     Completed Date:  1988-08-31     Revised Date:  2007-11-15    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  645-9     Citation Subset:  IM    
Affiliation:
Division of Cardiology, Cedars-Sinai Medical Center, Los Angeles, CA 90048.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Arrhythmias, Cardiac / etiology*
Electric Injuries / etiology*
Electric Power Supplies / adverse effects*
Electric Stimulation / instrumentation
Electrophysiology / instrumentation*
Equipment Safety
Humans
Grant Support
ID/Acronym/Agency:
HL07380/HL/NHLBI NIH HHS

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Sinus rhythm mapping in a canine model of ventricular tachycardia.
Next Document:  Torsade de pointes and other pause-induced ventricular tachycardias: the short-long-short sequence a...