Effects of heart rate, filling and slice thickness on the accuracy of left ventricular volume measurements in a dynamic cardiac phantom using ECG-gated MDCT.
MedLine Citation:
PMID:  18316343     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The objective of the study was to assess the effect of heart rate, filling condition and slice thickness on the accuracy of volumetric analysis based on multidetector-row computed tomography (MDCT) of a cardiac phantom. Retrospective electrocardiogram-gated MDCT of a pulsating phantom was performed under different conditions. End-diastolic volume (EDV) and end-systolic volume (ESV) for different heart rates (60-75 beats per minute), filling volumes and reconstructed slice thicknesses (2 mm and 5 mm) were obtained by three observers. Results were analysed by a linear mixed-effects model. Significant effects on the accuracy were found for heart rate (F-value, 7.3-39.2; p<0.004) and filling condition (F-value, 7.4-55.6; p<0.004), but not for slice thickness. Small relative differences in the assessment of EDV were found (range, -3% to 3%), but there was a trend for overestimation of the ESV (range, -1% to 18%). Underestimation of stroke volume and ejection fraction (range, -1% to -11%) became smaller under conditions of improved temporal resolution and larger EDV. Good interobserver agreement was found (SD <1.8 ml and <0.5%). In conclusion, MDCT allows sufficient and reliable measurements of ventricular volumes and calculation of left ventricle function for clinical applications. Heart rate and filling conditions significantly affect the accuracy of volumetrics, as demonstrated in this cardiac phantom. Thicker slices provide similar accuracy to thin slices.
Authors:
H Doğan; W J H Veldkamp; P Dibbets-Schneider; A M Spijkerboer; B J A Mertens; L J M Kroft; A De Roos; J Geleijns
Related Documents :
7235053 - Optimal control evaluation of left ventricular systolic dynamics.
10704173 - Improved mechanoenergetics and cardiac rest and reserve function of in vivo failing hea...
6654573 - The effect of modified fasting on blood pressure and sympathetic activity: a correlation?
Publication Detail:
Type:  Journal Article     Date:  2008-03-03
Journal Detail:
Title:  The British journal of radiology     Volume:  81     ISSN:  1748-880X     ISO Abbreviation:  Br J Radiol     Publication Date:  2008 Jul 
Date Detail:
Created Date:  2008-06-18     Completed Date:  2008-07-21     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0373125     Medline TA:  Br J Radiol     Country:  England    
Other Details:
Languages:  eng     Pagination:  577-82     Citation Subset:  AIM; IM    
Affiliation:
Department of Radiology, C2-S, Leiden University Medical Center, Albinusdreef 2, 2333 ZA Leiden, The Netherlands. h.dogan@lumc.nl
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Cardiac Volume / physiology*
Diastole
Electrocardiography / methods
Heart Rate / physiology*
Heart Ventricles / radiography
Humans
Observer Variation
Phantoms, Imaging
Systole
Tomography, X-Ray Computed / standards*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Robustified MANOVA with applications in detecting differentially expressed genes from oligonucleotid...
Next Document:  MRI of chronic sclerosing sialoadenitis.