Effects of Condensate in the Exhalation Limb of Neonatal Circuits on Airway Pressure During Bubble Continuous Positive Airway Pressure.
MedLine Citation:
PMID:  23481441     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
BackgroundBubble continuous positive airway pressure (B-CPAP) is frequently used in spontaneously breathing infants with lung disease. Often, the B-CPAP systems lack pressure alarms and pressure release valves. We observed a large volume of condensate in the exhalation limb of a patient circuit and conducted a series of experiments to test the hypothesis that accumulated condensate could affect delivered pressures.MethodsAn anatomically accurate nasal airway model of a preterm infant was attached to a spontaneously breathing lung model. A Fisher & Paykel B-CPAP system was attached to the nasal airway with bi-nasal short prongs and the rate of fluid condensation was measured. Next, tracheal pressures were monitored digitally to detect changes in airway pressure related to condensate accumulation. Measurements were obtained with volumes of 0, 5, 10, 15, and 20 mL of water in the exhalation limb at flow rates of 4, 6, 8, and 10 L/min. Measurements with 20 mL in the exhalation limb were recorded with and without an F&P pop-off valve in the circuit.ResultsThe rate of condensate accumulation was 3.8 mL/hour. At volumes of ≥10 mL, noticeable alterations in the airway pressure waveforms and significant increases in mean tracheal pressure were observed. The pop-off valve effectively attenuated peak tracheal pressures but only decreased mean pressures by 0.5-1.5 cmH2O.Discussion/ConclusionCondensate in the exhalation limb of the patient circuit during B-CPAP can significantly increase pressure delivered to the patient. The back and forth movement of this fluid causes oscillations in airway pressure that are much greater than the oscillations created by gas bubbling out the exhalation tube into the water bath. We recommend continuously monitoring pressure at the nasal airway interface, placing an adjustable pop-off valve in the circuit set to 5 cmH2O above desired mean pressure, and emptying fluid from the exhalation limb every 2-3 hours.
Authors:
Tiffany M Youngquist; C Peter Richardson; Robert M Diblasi
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2013-3-12
Journal Detail:
Title:  Respiratory care     Volume:  -     ISSN:  0020-1324     ISO Abbreviation:  Respir Care     Publication Date:  2013 Mar 
Date Detail:
Created Date:  2013-3-13     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7510357     Medline TA:  Respir Care     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
Center for Developmental Therapeutics, Seattle Children's Research Institute, Seattle, WA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Pulmonary Function and Flow Volume Loop Patterns in Patients with Tracheobronchomalacia.
Next Document:  Status of non-standard neutrino interactions.