Document Detail


Effect of measurement conditions on measured levels of peak exhaled nitric oxide.
MedLine Citation:
PMID:  9337828     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: It is possible to measure nitric oxide (NO) levels in exhaled air. The absolute concentrations of exhaled NO obtained by separate workers in similar patient groups and normal subjects with apparently similar techniques have been very different. A study was undertaken to determine whether changes in measurement conditions alter the concentration of exhaled NO. METHOD: NO concentrations measured by a chemiluminescence analyser (Dasibi Environmental Corporation) and carbon dioxide (CO2) measured by a Morgan capnograph were analysed in single exhalations from total lung capacity in healthy volunteers (mean age 35.9 years). Ten subjects performed five exhalations at four different expiratory flow rates, at four different expiratory mouth pressures, and before and after drinking hot (n = 5) or cold (n = 5) water. Three subjects performed five exhalations on a day of high background NO (mean NO level 134 ppb) before and after a set of five exhalations made while both the subject and analysers were sampling from a low NO/NO-free reservoir system. RESULTS: The mean peak concentration of NO decreased by 35 ppb (95% CI 25.7 to 43.4) from a mean (SE) of 79.0 (15.5) ppb at an expiratory flow rate of 250 ml/min to 54.1 (10.7) ppb at 1100 ml/min. The mean peak concentration of NO did not change significantly with change in mouth pressure. The mean (SE) peak NO concentration decreased from 94.4 (20.8) ppb to 70.8 (16.5) ppb (p = 0.002, 95% CI 12.9 to 33.1) with water consumption. The mean NO concentration with machine and subject sampling from the low NO reservoir was 123.1 (19.4) ppb, an increase from results obtained before (81.9 (10.2) ppb, p = 0.001, 95% CI -19.9 to -62.7) and after (94.2 (18.3) ppb, p = 0.017, 95% CI 6.0 to 51.8) sampling with high ambient NO. CONCLUSIONS: The measurement of exhaled NO must be performed in a carefully standardised manner to enable different teams of investigators to compare results.
Authors:
C A Byrnes; S Dinarevic; C A Busst; E A Shinebourne; A Bush
Related Documents :
1387738 - A safe and simple method to maintain a clear field of vision during laparoscopic cholec...
17269878 - Comparison of noninvasive cardiac output measured by use of partial carbon dioxide rebr...
6800988 - Regulation of paco2 and ventilation in humans inspiring low levels of co2.
14551718 - Does high-pressure carbon dioxide insufflation facilitate mucosal dissection in transan...
3999698 - Cardiac performance of isolated beating hearts obtained from rats anesthetized by three...
23859938 - Cardiovascular coupling analysis with high-resolution joint symbolic dynamics in patien...
Publication Detail:
Type:  Clinical Trial; Controlled Clinical Trial; Journal Article    
Journal Detail:
Title:  Thorax     Volume:  52     ISSN:  0040-6376     ISO Abbreviation:  Thorax     Publication Date:  1997 Aug 
Date Detail:
Created Date:  1997-11-04     Completed Date:  1997-11-04     Revised Date:  2009-11-18    
Medline Journal Info:
Nlm Unique ID:  0417353     Medline TA:  Thorax     Country:  ENGLAND    
Other Details:
Languages:  eng     Pagination:  697-701     Citation Subset:  IM    
Affiliation:
Paediatric Department, Royal Brompton Hospital, London, UK.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Adult
Analysis of Variance
Breath Tests / methods*
Capnography
Chemiluminescent Measurements
Drinking
Female
Humans
Male
Middle Aged
Nitric Oxide / analysis*
Peak Expiratory Flow Rate
Sensitivity and Specificity
Chemical
Reg. No./Substance:
10102-43-9/Nitric Oxide
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Physiological effects and optimisation of nasal assist-control ventilation for patients with chronic...
Next Document:  Optimal treatment of descending necrotising mediastinitis.