| Effect of low concentrations of halothane on the oxygen electrode. | |
| | |
MedLine Citation:
|
PMID: 7426229 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
The effect of low concentrations of halothane dissolved in saline (equivalent to 0.34% halothane) on the oxygen electrode was examined using a multi-wire Clark type electrode, consisting of eight platinum wires, each of 15 microns tip diameter. On the first exposure of the day to saline equilibrated with halothane in 10% oxygen or air, the output of the oxygen electrode increased by 40% over 56 min. On the second and subsequent exposures the increase in output was much less, but was still 10% greater than the correct values. After immersing the electrode overnight in saline without halothane the high sensitivity to halothane was present again next morning, even if neither membrane nor electrolyte was changed. |
| | |
Authors:
|
T Maekawa; Y Okuda; D G McDowall |
Related Documents
:
|
851499 - Cardiovascular effects of gallamine triethiodide and succinylcholine chloride during ha... 1928769 - Physiologic responses to mild perianesthetic hypothermia in humans. 337989 - A comparison of the effect of three anaesthetic techniques on postoperative arterial ox... 636809 - Hemodynamics during superficial thiopental anesthesia in the dog. 16874309 - Bioequivalence of topical formulations in humans: evaluation by dermal microdialysis sa... 15787919 - Syringe pumps take too long to give occlusion alarm. 17043849 - Workplace no and no2 during combined treatment of infants with nasal cpap and no. 20833639 - Exercise therapy in polycystic ovary syndrome: a systematic review. 10772799 - Antagonizing activity of chick grg4 against tectum-organizing activity. |
Publication Detail:
|
Type: Journal Article; Research Support, Non-U.S. Gov't |
Journal Detail:
|
Title: British journal of anaesthesia Volume: 52 ISSN: 0007-0912 ISO Abbreviation: Br J Anaesth Publication Date: 1980 Jun |
Date Detail:
|
Created Date: 1981-01-26 Completed Date: 1981-01-26 Revised Date: 2006-11-15 |
Medline Journal Info:
|
Nlm Unique ID: 0372541 Medline TA: Br J Anaesth Country: ENGLAND |
Other Details:
|
Languages: eng Pagination: 585-7 Citation Subset: IM |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Blood Gas Analysis
/
instrumentation,
methods* Halothane* Methoxyflurane Microelectrodes* Platinum Time Factors |
| Chemical | |
Reg. No./Substance:
|
151-67-7/Halothane; 7440-06-4/Platinum; 76-38-0/Methoxyflurane |