The Effect of Ventilator Performance on Airway Pressure Release Ventilation: A Model Lung Study.
MedLine Citation:
PMID:  21519042     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
Background: Using a model lung connected to six different ventilators, with each ventilator in the airway pressure release ventilation mode, we measured differences in intrinsic positive end-expiratory pressure (PEEPi) during the expiratory phase and calculated the inspiratory and expiratory pressure time product (PTP) as an index of work of breathing during the inspiratory phase. Methods: We compared 6 ventilators: Puritan-Bennett 840, Evita XL, Servo i, Avea, Hamilton G5, and Engström. With a constant inspiratory pressure level of 25 cm H(2)O and expiratory pressure level of 0 cm H(2)O, PEEPi was measured as the expiratory time was decremented from 1.0 second to 0.2 second in steps of 0.1 second. The inspiratory and expiratory PTPs were measured during the ventilator's inspiratory phase by simulating spontaneous breathing with a tidal volume of 300 mL, with a respiratory rate of 30 breaths/min and with expiratory flow rates of 0.5 L/s, 1.0 L/s, and 1.5 L/s. Results: In all ventilators, the progressive diminution of the expiratory time caused a significant increase in PEEPi (P < 0.001). With a 0.2-second expiratory time, PEEPi ranged from 9.4 ± 0.07 cm H(2)O for the Servo i to 15.7 ± 0.04 cm H(2)O for the Avea. The Servo i had a significantly lower inspiratory PTP than did the other ventilators (P < 0.001). When the expiratory flow rate was 0.5 L/s and 1.0 L/s, the expiratory PTP was lower with the Servo i and Evita XL than with the other ventilators (P < 0.001). Conclusions: PEEPi varied significantly among ventilators. Inspiratory and expiratory work of breathing varied between ventilators when spontaneous breathing occurred during the ventilator's inspiratory phase.
Authors:
Takeshi Yoshida; Akinori Uchiyama; Takashi Mashimo; Yuji Fujino
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2011-4-25
Journal Detail:
Title:  Anesthesia and analgesia     Volume:  -     ISSN:  1526-7598     ISO Abbreviation:  -     Publication Date:  2011 Apr 
Date Detail:
Created Date:  2011-4-26     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  1310650     Medline TA:  Anesth Analg     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
From the Department of Anesthesiology and Intensive Care Medicine, Osaka University Graduate School of Medicine, Suita, Japan.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Mixed germ cell tumor with embryonal carcinoma, choriocarcinoma, and epithelioid trophoblastic tumor...
Next Document:  Multicenter Assessment of the Iowa Satisfaction with Anesthesia Scale, an Instrument that Measures P...