Dual energy x-ray absorptiometry: the effects of beam hardening on bone density measurements.
MedLine Citation:
PMID:  1584146     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
X-ray tubes have superseded radionuclide sources for dual photon absorptiometry of the spine and hip. However, the use of a polyenergetic spectrum is a potential source of error for x-ray absorptiometers since beam hardening may result in a nonlinear measurement scale for bone mineral density (BMD). A quantitative study of the effects of beam hardening on measurements made with a commercial dual energy x-ray scanner has been performed. Bone was represented by layers of aluminum of linearly increasing thickness which were scanned under water thicknesses ranging from 0 to 25 cm to represent different body thicknesses of soft tissue. Beam hardening had two effects on measured BMD: (i) at a constant true BMD, measured BMD varied with water thickness; (ii) at a constant water thickness, the BMD scale was not precisely linear. For conditions appropriate to spine and hip studies (BMD) values in the range 0.7 to 1.4 g/cm2 and body thickness between 15 and 25 cm) the maximum deviation of measured BMD from a linear scale was 0.023 g/cm2, while the root-mean-square deviation (0.01 g/cm2) was comparable to the measurement precision for a spine or femoral neck scan (about 1%). The largest departures from linearity were found to occur at the thinnest water thicknesses for BMD values in the range 0.2 to 0.6 g/cm2. The effect of scale nonlinearity on the results of longitudinal studies was examined: for a spine scan at 20-cm body thickness, measured changes in BMD slightly overestimated the true change and implied an error of 0.15%/year for a measurement of a true rate of loss of 3% year in a postmenopausal woman.
Authors:
G M Blake; D B McKeeney; S C Chhaya; P J Ryan; I Fogelman
Related Documents :
2887716 - Effects of h1 blocking agents on age-related osteopenia in the mouse.
774996 - Genesis of the sweat:plasma urea concentration gradient.
573346 - Body composition changes in the critically ill patient: emphasis on water balance.
7156526 - Respiration and metabolism in the giraffe.
18842056 - Molecular dynamics simulation of the cooperative adsorption of barley lipid transfer pr...
23087696 - Experimental vs. modeled water use in mature norway spruce (picea abies) exposed to ele...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Medical physics     Volume:  19     ISSN:  0094-2405     ISO Abbreviation:  Med Phys     Publication Date:    1992 Mar-Apr
Date Detail:
Created Date:  1992-06-12     Completed Date:  1992-06-12     Revised Date:  2007-11-15    
Medline Journal Info:
Nlm Unique ID:  0425746     Medline TA:  Med Phys     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  459-65     Citation Subset:  IM    
Affiliation:
Department of Nuclear Medicine, Guy's Hospital, London, United Kingdom.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Absorptiometry, Photon*
Bone Density*
Humans
Models, Structural

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Imaging characteristics of different mammographic screens.
Next Document:  Dynamic pulmonary imaging: performance properties of a digital fluoroscopy system.