Dosimetry concepts for scanner quality assurance and tissue dose assessment in micro-CT.
MedLine Citation:
PMID:  22320775     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
Purpose: At present, no established methods exist for dosimetry in micro computed tomography (micro-CT). The purpose of this study was therefore to investigate practical concepts for both dosimetric scanner quality assurance and tissue dose assessment for micro-CT.Methods: The computed tomography dose index (CTDI) was adapted to micro-CT and measurements of the CTDI both free in air and in the center of cylindrical polymethyl methacrylate (PMMA) phantoms of 20 and 32 mm diameter were performed in a 6 month interval with a 100 mm pencil ionization chamber calibrated for low tube voltages. For tissue dose assessment, z-profile measurements using thermoluminescence dosimeters (TLDs) were performed and both profile and CTDI measurements were compared to Monte Carlo (MC) dose calculations to validate an existing MC tool for use in micro-CT. The consistency of MC calculations and TLD measurements was further investigated in two mice cadavers.Results: CTDI was found to be a reproducible quantity for constancy tests on the micro-CT system under study, showing a linear dependence on tube voltage and being by definition proportional to mAs setting and z-collimation. The CTDI measured free in air showed larger systematic deviations after the 6 month interval compared to the CTDI measured in PMMA phantoms. MC calculations were found to match CTDI measurements within 3% when using x-ray spectra measured at our micro-CT installation and better than 10% when using x-ray spectra calculated from semi-empirical models. Visual inspection revealed good agreement for all z-profiles. The consistency of MC calculations and TLD measurements in mice was found to be better than 10% with a mean deviation of 4.5%.Conclusions: Our results show the CTDI implemented for micro-CT to be a promising candidate for dosimetric quality assurance measurements as it linearly reflects changes in tube voltage, mAs setting, and collimation used during the scan, encouraging further studies on a variety of systems. For tissue dose assessment, MC calculations offer an accurate and fast alternative to TLD measurements allowing for dose calculations specific to any geometry and scan protocol.
Authors:
Martin Hupfer; Daniel Kolditz; Tristan Nowak; Fabian Eisa; Robert Brauweiler; Willi A Kalender
Related Documents :
8475815 - Lipoma of the cerebral cortex: case report.
10372845 - Primary antiphospholipid syndrome and cerebral atrophy: a rare association?
8686435 - Correlation between cerebral blood flow values obtained by xenon/ct and kety-schmidt (n...
9568925 - Acute childhood hemiplegia associated with chickenpox.
10074725 - 133xenon ventilation scan as a functional assessment in bronchiectasis.
20203355 - Spin-glass-like freezing and enhanced magnetization in ultra-small cofe2o4 nanoparticles.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Medical physics     Volume:  39     ISSN:  0094-2405     ISO Abbreviation:  Med Phys     Publication Date:  2012 Feb 
Date Detail:
Created Date:  2012-02-10     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0425746     Medline TA:  Med Phys     Country:  United States    
Other Details:
Languages:  eng     Pagination:  658     Citation Subset:  IM    
Affiliation:
Institute of Medical Physics, University of Erlangen-Nürnberg, Henkestraße 91, 91052 Erlangen, Germany.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  XCOM intrinsic dimensionality for low-Z elements at diagnostic energies.
Next Document:  Correction method for in-air output ratio for output variations occurring with changes in backscatte...