Document Detail


Does an SVC electrode further reduce DFT in a hot-can ICD system?
MedLine Citation:
PMID:  9121982     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Pectorally implanted ICDs that defibrillate with the RV electrode and the ICD housing have gained clinical acceptance. However, it is still debatable whether adding an SVC electrode connected to the housing will further reduce the threshold of defibrillation (DFT). This study utilized eight pigs. DFTs were measured with a 50 V step-down protocol starting at 650 V (20 J). Shock strength for 50% success (E50) was estimated with the average of three reversals. In addition to alpha dummy device, Lead I (Pacesetter Models 1558 and 1538) or Lead II (Endotak 72) were used. Leads I are active fixation, true bipolar sensing with 5-cm shocking coils. Lead II has an integrated bipolar sensing with a 4.7-cm RV and 6.9-cm SVC shocking coils. A 95 microF defibrillation system was used to deliver a 44% tilt tuned biphasic 1.6/2.5 ms waveform, and to measure lead impedance. The RV electrode was the anode during phase I. With Lead IRV-->CAN the DFT was 531 +/- 75 V (13.6 +/- 3.8J) and the E50 was 496 +/- 89 (12 +/- 4.3 J). These were not significantly (NS) different than the DFT for RV-->CAN and SVC which was 518 +/- 84 V (13 +/- 4.2 J) or the E50 which was 476 +/- 84 V (11 +/- 3.9 J). Similar results were obtained with Lead II. Despite a decrease in lead impedance there was no apparent benefit from the addition of the SVC electrode. Lead I provided equivalent DFT performance to Lead II.
Authors:
G A Mouchawar; W K Wolsleger; P D Doan; J D Causey; M W Kroll
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  20     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  1997 Jan 
Date Detail:
Created Date:  1997-04-23     Completed Date:  1997-04-23     Revised Date:  2003-11-14    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  163-7     Citation Subset:  IM    
Affiliation:
Pacesetter, Inc., Sylmar, California 91392-9221, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Analysis of Variance
Animals
Chronaxy
Defibrillators, Implantable*
Electric Conductivity
Electric Countershock / methods*
Electric Impedance
Electric Stimulation
Electrocardiography
Electrodes, Implanted*
Equipment Design
Heart Ventricles
Swine
Vena Cava, Superior*
Ventricular Fibrillation / therapy

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Comparison of defibrillation efficacy using biphasic waveforms delivered from various capacitances/p...
Next Document:  Temporary transvenous cardioversion and defibrillation: a new method for practical tachyarrhythmia m...