Document Detail


Dislodgement of atrial lead of the CRT-D system and its migration into the pocket presenting as its absence on chest x-ray.
MedLine Citation:
PMID:  17437589     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Lead dislodgement and migration is a known complication after implantation of cardiac rhythm devices. We present a case of dislodgement of atrial lead of the CRT-D system and its migration into the device pocket. The device itself had rotated by 90 degrees but the right and left ventricular leads were not at all affected. The case is unique because the lead had coiled up and was lying behind the device in the pocket and hence was not even seen on a chest x-ray.
Authors:
Kartikeya Bhargava; Tarlochan Singh Kler
Related Documents :
16878609 - Clinical experience with the tandemheart percutaneous ventricular assist device.
3548779 - Temporary ventricular support. current indications and results.
8258439 - Selection of myocardial electrogram features for use by implantable devices.
11833709 - "single suture" for exposure of the heart in left ventricular assist device placement.
17489349 - Treatment of a distal left main trifurcation supported by the tandemheart left ventricu...
8944979 - Mechanical circulatory support for post cardiotomy cardiogenic shock in infants.
8223729 - Radionuclide assessment of regional differences in left ventricular wall motion and myo...
9149549 - Cardiac fibroma presenting as sudden death in a six-month-old infant.
20588189 - A randomized invasive cardiac electrophysiology study of the combined ion channel block...
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  30     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  2007 Apr 
Date Detail:
Created Date:  2007-04-17     Completed Date:  2007-06-28     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  584-5     Citation Subset:  IM    
Affiliation:
Escorts Heart Institute and Research Centre, New Delhi, India. drkartikeya@hotmail.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Bundle-Branch Block / physiopathology,  therapy*
Defibrillators, Implantable / adverse effects*
Electrocardiography
Equipment Failure
Foreign-Body Migration / radiography*
Humans
Male
Radiography, Thoracic

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Value of pacemaker atrial diagnostic data in patients with paroxysmal atrial fibrillation: an opport...
Next Document:  The influence of anaemia on stroke prognosis and its relation to N-terminal pro-brain natriuretic pe...