| Differences in oscillometric non-invasive blood pressure measurements recorded by different revisions of the Philips Component Monitoring System. | |
| | |
MedLine Citation:
|
PMID: 16077268 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
OBJECTIVE: To investigate whether a simulator can detect differences in oscillometric noninvasive blood pressure measurements in different revisions of the Philips Component Monitoring System. Philips revised the intraarterial-based algorithm with software revision C.0, and added an auscultation-based algorithm in revision C.1. METHODS: Three methods were adopted. Firstly, differences were assessed using the Dynatech Nevada CuffLink simulator. Five noninvasive blood pressure modules were tested on monitors with software revision C (software revision prior to C.0) and revision C.0. Monitor C.1 allows the selection of the original intraarterial algorithm (as for revision C), the improved intraarterial (revision C.0) or auscultation; two noninvasive blood pressure modules were tested on it.Secondly, pressures were recorded simultaneously from a participant's left and right arms by monitors C and C.0. The monitors were then swapped between arms and a second set of six pressures were recorded. The two sets were subtracted to calculate the differences between monitors, removing interarm differences. The experiment was carried out on two participants.Thirdly, pressures were recorded from three participants with a simultaneous reference auscultatory measurement. Six sets of simultaneous oscillometric and auscultatory determinations were recorded from monitors C and C.0. RESULTS AND CONCLUSIONS: All three methods revealed that the 'original' intraarterial-based algorithm in revision C recorded lower systolic, diastolic and mean arterial pressures than revision C.0, with the difference in diastolic pressure being the greatest. The difference in diastolic pressures between C and C.0 was up to 15 mmHg. The simulator evaluation finding of higher systolic and diastolic pressures recorded by the revised intraarterial algorithm (C.1 option 3) than by the auscultation-based algorithm (C.1 option 2) is in qualitative agreement with the regression equations of the clinical validation studies. The results suggest that simulators can identify the existence of differing algorithms used to measure blood pressure by the CMS monitor. Philips has continued to develop the algorithms used to determine the pressure by the oscillometric method. While significantly different pressures were recorded by the different revisions, it is interesting that each revision passed clinical validation studies. |
| | |
Authors:
|
Ian C Murray; John N Amoore; David H T Scott |
Related Documents
:
|
7995638 - Systematic difference between blood pressure readings caused by cuff type. 9434918 - Non-invasive measurement of cardiac output: whole-body impedance cardiography in simult... 19275828 - Use of evidence in hypertension guidelines: should we measure in both arms? 18755068 - Accuracy of automated blood pressure monitors. 15037878 - Should pulse pressure become part of the framingham risk score? 11751728 - Pulse pressure, endothelium function, and arterial stiffness in spontaneously hypertens... |
Publication Detail:
|
Type: Clinical Trial; Journal Article; Research Support, Non-U.S. Gov't |
Journal Detail:
|
Title: Blood pressure monitoring Volume: 10 ISSN: 1359-5237 ISO Abbreviation: Blood Press Monit Publication Date: 2005 Aug |
Date Detail:
|
Created Date: 2005-08-03 Completed Date: 2005-10-25 Revised Date: 2006-11-15 |
Medline Journal Info:
|
Nlm Unique ID: 9606438 Medline TA: Blood Press Monit Country: England |
Other Details:
|
Languages: eng Pagination: 215-22 Citation Subset: IM |
Affiliation:
|
Department of Medical Physics, Royal Infirmary of Edinburgh, Edinburgh, UK. I.C.Murray-2@sms.ed.ac.uk |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Algorithms Auscultation Blood Pressure Blood Pressure Determination / instrumentation*, methods* Computer Simulation Humans Monitoring, Physiologic / instrumentation*, methods* Periodicity Software* |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine