Document Detail

Deviation between navigated and final 3-dimensional implant position in mini-invasive unicompartmental knee arthroplasty: a pilot study in 13 patients.
Jump to Full Text
MedLine Citation:
PMID:  23043273     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND AND PURPOSE: Unicompartmental knee arthroplasty (UKA) is an established method of treating isolated gonartrosis. Modern techniques such as computer-assisted surgery (CAS) and minimally invasive surgery (MIS) are attractive complementary methods to UKA. However, the positioning of the components remains a concern. Thus, we performed a prospective study to assess whether there was deviation between the navigated implant position and the final implant position.
PATIENTS AND METHODS: We performed UKA with MIS and CAS in 13 patients. By comparing intraoperative navigation data with postoperative computed tomography (CT) measurements, we calculated the deviation between the computer-assisted implant position and the final 3-D implant position of the femoral and tibial components.
RESULTS: The computer-assisted placement of the femoral and tibial component showed adequate position and consistent results regarding flexion-extension and varus-valgus. However, regarding rotation there was a large variation and 6 of 10 patients were outside the target range for both the femoral component and the tibial component.
INTERPRETATION: Difficulties in assessing anatomical landmarks with the CAS in combination with MIS might be a reason for the poor rotational alignment of the components.
Authors:
Nicolas Martinez-Carranza; Lars Weidenhielm; Joakim Crafoord; Margareta Hedström
Related Documents :
24013393 - Primary stability determination of implants inserted in sinus augmented sites: 1-step v...
1317993 - Tissue breakdown and exposure associated with orbital hydroxyapatite implants.
12078593 - Immediate restoration of multiple tooth implants for aesthetic implant restorations.
Publication Detail:
Type:  Comparative Study; Journal Article; Research Support, Non-U.S. Gov't     Date:  2012-10-08
Journal Detail:
Title:  Acta orthopaedica     Volume:  83     ISSN:  1745-3682     ISO Abbreviation:  Acta Orthop     Publication Date:  2012 Dec 
Date Detail:
Created Date:  2012-11-26     Completed Date:  2013-01-24     Revised Date:  2013-07-11    
Medline Journal Info:
Nlm Unique ID:  101231512     Medline TA:  Acta Orthop     Country:  England    
Other Details:
Languages:  eng     Pagination:  625-8     Citation Subset:  IM    
Affiliation:
Department of Orthopaedics, Karolinska University Hospital, Stockholm, Sweden. nicolas.martinez-carranza@karolinska.se
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Arthroplasty, Replacement, Knee / adverse effects,  methods*
Female
Follow-Up Studies
Humans
Imaging, Three-Dimensional*
Knee Prosthesis*
Male
Middle Aged
Pilot Projects
Prospective Studies
Prosthesis Failure
Prosthesis Fitting / instrumentation,  methods
Risk Assessment
Sampling Studies
Surgery, Computer-Assisted / adverse effects,  methods*
Surgical Procedures, Minimally Invasive / adverse effects,  methods
Time Factors
Tomography, X-Ray Computed / methods
Treatment Outcome
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): Acta Orthop
Journal ID (iso-abbrev): Acta Orthop
Journal ID (publisher-id): ORT
ISSN: 1745-3674
ISSN: 1745-3682
Publisher: Informa Healthcare
Article Information
Download PDF
Copyright: © Nordic Orthopaedic Federation
open-access:
Received Day: 15 Month: 6 Year: 2012
Accepted Day: 15 Month: 7 Year: 2012
Print publication date: Month: 12 Year: 2012
Electronic publication date: Day: 26 Month: 11 Year: 2012
Volume: 83 Issue: 6
First Page: 625 Last Page: 628
PubMed Id: 23043273
ID: 3555452
DOI: 10.3109/17453674.2012.736840
Publisher Id: ORT_A_736840_O

Deviation between navigated and final 3-dimensional implant position in mini-invasive unicompartmental knee arthroplasty : A pilot study in 13 patients
Nicolas Martinez-Carranza12
Lars Weidenhielm13
Joakim Crafoord4
Margareta Hedström12
1Department of Orthopaedics, Karolinska University Hospital
2Institution of Clinical Sciences, Intervention and Technology (CLINTEC), Karolinska Institute
3Department of Molecular Medicine and Surgery, Karolinska Institute
4Department of Radiology, Karolinska University Hospital, Stockholm, Sweden
Correspondence: Correspondence: nicolas.martinez-carranza@karolinska.se

The advantages of unicompartmental knee arthroplasty (UKA) such as less blood loss, shorter hospital stay, shorter time to recovery, superior kinematics, better range of motion, and bone stock preserving capacity are well documented (Saccomanni 2010). On the other hand, UKA has a longer learning curve and surgeons usually perform fewer surgeries a year; it is therefore associated with a higher risk of complications such as component malpositioning performed with minimally invasive (MIS) technique (Lindstrand et al. 2000, Romanowski and Repicci 2002).

UKA has been controversial since its introduction in the 1970s (Deshmukh and Scott 2001), when polyethylene wear (Collier et al. 1991) and loosening—partially due to suboptimal component positioning (Moreland 1988, Bathis et al. 2004)—were the predominant failure mechanisms. However, recent reports have shown more promising results (Hollinghurst et al. 2006, Griffin et al. 2007), and there has been renewed interest in UKA in recent years (Bert 2005). Part of this can be explained by improved component design and surgical technique, but MIS (Laskin 2001) especially may have had a key role in this trend. Proper and consistent implant positioning is vital, and computer-assisted surgery (CAS) has been suggested to improve the accuracy of implant positioning (Rosenberger et al. 2008). Few studies have assessed the implant positioning of navigated UKAs operated with MIS technique (Perlick et al. 2004, Kim et al. 2005, Konyves et al. 2010) and, as far as we know, none have assessed it with three-dimensional (3-D) CT.

We determined the degree of 3-D deviation between the navigated implant position and the final implant position in UKA performed with MIS technique.


Patients and methods

In a prospective pilot study, 13 consecutive patients (7 women) with a median age of 65 (50–76) years underwent cemented primary UKA (Preservation; Depuy Inc, Warsaw, IN) using a computer-assisted navigation system (VectorVision; BrainLab, Munich, Germany) in combination with a minimally invasive incision. 2 senior surgeons (MH and LW) performed all surgeries.

All patients met the inclusion criteria: radiographic medial osteoarthrosis (OA), intact cruciate ligaments, flexion contracture of less than 10 degrees, varus deformity of less than 10 degrees, and symptoms of medial arthritis that warranted joint replacement. 2 patients were excluded: in 1 patient, the tracker loosened during surgery and was thus not navigated; in the other patient, only the tibia was measured peroperatively due to technical difficulties with the CAS software program. In one patient, the rotational component position was not measured postoperatively.

The navigation system used was an image-free instrumentation tool for determination of the component position and mechanical alignment intraoperatively by localizing the center of the hip, knee, and ankle.

Computer-assisted surgery

After exposing the knee through a medial parapateller arthrotomy of not more than 8 cm (Repici and Eberle 1999), 2 passive trackers were rigidly fixed to the femur and tibia and the registration process was carried out (Figure 1). The hip was rotated to determine the center of the hip. Specific landmarks of the knee were digitized to determine the center of the knee, and these points were marked with 1.0-mm tantalum balls in the femur and in the tibia, inserted with an injection gun specially made for this purpose. They were placed in the subchondral bone, marking the location of the point that we registered with our Ci pointer in the navigation system as the center of the tibial plateau and the center of the distal femur. The center of the distal femur was identified as a point slightly medial to the anterior aspect of the femoral notch and the center of the proximal tibia was identified as the posterior aspect of the ACL insertion on the tibial plateau according to the manual for the Ci navigation system. Since the definition of these points is somewhat imprecise, we chose to mark this point with a tantalum ball in order to be able to identify and use the same point when we made our postoperative measurements at the postoperative CT examination. The rotation of the femur was referenced from the transepicondylar axis. The landmarks used to determine the rotation of the tibia were the AP direction of the tibia and the most anterior, medial, and posterior aspects of the medial tibial plateau. The lateral and medial malleolus were localized to define the center of the talus. The computer created a virtual model from the supplemented registration data. The orientation of the cutting blocks was determined using the navigation system (Figure 2), and after bone resection the cutting planes were documented.

Computed tomography measurements

A CT scan was performed in each patient postoperatively. The scans were made on a General Electric Volume CT (64-slice CT), with a slice thickness of 0.625 mm, 120 kV, 200 mA, 0.6-s rotation time, a collimation width of 40 mm, and a spiral pitch factor of 0.98. The images were then reconstructed with a “bone” algorithm in three planes.

The scan included 3 anatomical regions: (1) the femoral head, (2) the knee, and (3) the ankle. A radiologist (JC) defined the following anatomical landmarks: center of the hip, knee, ankle, and posterior femoral condylar plane. Femoral implant landmarks (antero-superior apex, distal peak, and postero-superior corner) and tibial implant landmarks (anterior-inferior corner, medial-inferior border, and posterior-inferior corner) were used. The position of the femoral component was calculated using the center of the hip, knee, and ankle, the posterior femoral condylar plane, and landmarks on the femoral implant. Subsequently, position of the the tibial component was calculated using the center of the hip, knee, and ankle, the posterior femoral condylar plane, and landmarks on the tibial implant.

Varus/valgus, flexion/extension, and external/internal rotation of both the femoral and tibial components were assessed and expressed as average values. We used the anatomical transepicondylar line to determine the rotation of the femoral component intraoperatively with CAS, and measured the final femoral implant position on postoperative CT scan using the posterior condylar plane.

The average deviation between the intraoperative navigation value and the actual postoperative implant position as measured with CT scan was calculated. We counted how many patients were outside (in any direction) of an arbitrary target range of 3 degrees from the desired original intraoperative CAS measurement. Although this value is an arbitrarily defined safety zone for satisfactory implant position, it is based on theoretical considerations—as errors in CAS, movements of the cutting block/saw, and the cementing of components contribute to differences in peroperative and postoperative measurements (Catani et al. 2008).

Ethics

All patients agreed to participate in the trial and ethical approval was obtained.


Results

Of the 11 patients, component rotation could only be assessed in 10 patients due to loss of component rotation measurement in one patient. All patients showed similar alignment values for the components in varus-valgus and flexion-extension when comparing peroperative component position (as measured with the aid of navigation) with postoperative CT measurements. Regarding the rotation of the components, there was great variation in the positioning and a large difference between peroperative and postoperative measurements. 8 of 10 patients had a rotational deviation of the femoral component of more than 3 degrees, and 6 of 10 patients had a rotational deviation of the tibial component of more than 3 degrees (Tables 1 and 2).


Discussion

The importance of proper component positioning for the long-term survival of TKR is well known. In TKR, a threshold of 3 degrees of deviation in the coronal plane has been reported to increase the risk of loosening (Mason et al. 2007). Some reports have shown the significance of accurate alignment in varus-valgus alignment in relation to the survivorship of the prosthesis in UKA (Weale et al. 2000, Deshmukh and Scott 2001) where 50% of the cases requiring revision were due to overcorrection (Barett and Scott 1987, Chakrabarty et al. 1998). To the best of our knowledge, there have been no previous reports on navigated MIS UKA in which the component positioning has been determined postoperatively with 3-D CT scan.

We found a large deviation between intraoperative CAS and postoperative CT scan measurements for the rotational position of both the tibial and the femoral component. A similar result was found in a recent study assessing the tibial component rotation in UKR using CT scan (Servien et al. 2011). One reason for the large variation in the rotational plane in MIS UKA could be the difficulty in identifying landmarks. The epicondyles and Whiteside’s line are not as visible as in a standard approach and are therefore less reliable. A recent report on TKR showed a considerable difference in the rotational plane between the intraoperative CAS component position and postoperative CT measurements (van der Linden-van der Zwaag et al. 2011). With our CAS system, it is not possible to measure the final position of the components after cementing. Nonetheless, the malpositioning was only found in the rotational plane and not in the 2 other directions and was therefore probably not related solely to the cementing technique. It is possible that improving the navigation system we used in this study, or using another system, might have yielded better results.

Overall, little has been reported regarding rotational alignment, even though it plays an important role in knee kinematics (Schnurr et al. 2009). The landmarks in the coronal plane are clinically and radiographically well defined and the long-term effects of malpositioning of the UKA components in terms of survival are well documented (Ridgeway et al. 2002). However, the landmarks for rotational UKA component positioning are less well defined and the clinical significance of malrotation is not fully understood—although it has been suggested as a major cause of pain in TKR (Nicoll and Rowley 2010).

The advantage of UKA might be enhanced with MIS technique. According to the literature, the rationale for this technique is less muscular damage, resulting in faster rehabilitation and less pain. We used the MIS technique because it is frequently used in UKA, although this might stress the surgeon’s ability and thus increase the risk of component malpositioning, compromising the long-term result. In the present study, MIS UKA performed with CAS had the advantages of a less invasive surgery with no component malpositioning, in varus-valgus and flexion-extension. We do not know whether a standard approach would have resulted in better component positioning, but comparison between MIS and a standard surgical approach was not within the scope of this study.

One weakness of our pilot study was that it was underpowered and lacked a control group. The advantage was that fewer patients were put at risk if the results were not encouraging. Our intention was to conduct a randomized controlled study comparing conventional and navigated technique, but the results of this pilot study have not been promising enough to encourage us to proceed with it.


NM-C: writing of the manuscript; JC: writing of CT method and CT measurements; LW and MH: study design and performance of all surgeries. All authors: data analysis and review of the manuscript.

We thank Depuy, Johnson and Johnson, Sweden for financial support that enabled us to perform the CT measurements.

No competing interests declared.


References
Barett WP,Scott RD. Revision of failed unicondylar knee arthroplastyJ. Bone Joint Surg (Br)Year: 1987691328
Bathis H,Perlick L,Tingart M,Luring C,Zurakowski D,Grifka J. Alignment in total knee arthroplasty. A comparison of computer-assisted surgery with the conventional techniqueJ Bone Joint Surg (Br)Year: 200486682715274263
Bert JM. Unicompartmental knee replacementOrthop Clin North AmYear: 2005365132216164956
Catani F,Biasca N,Ensini A,Leardini A,Bianchi L,Digennaro V,Giannini S. Alignment deviation between bone resection and final positioning in computer-navigated total knee arthroplastyJ Bone Joint Surg (Am)Year: 20089047657118381314
Chakrabarty G,Newman JH,Ackroyd CE. Revision of unicompartmental arthroplasty. Clinical and technical considerationsJ. ArthroplastyYear: 1998131919526213
Collier JP,Mayor MB,McNamara JL,Surprenant VA,Jensen RE. Analysis of the failure of 122 polyethylene inserts from uncemented tibial knee componentsClin OrthopYear: 1991273232421959276
Deshmukh RV,Scott RD. Unicompartmental knee arthroplasty: long-term resultsClin OrthopYear: 2001392272811716395
Griffin T,Rowden N,Morgan D,Atkinson R,Woodruff P,Maddern G. Unicompartmental knee arthroplasty for the treatment of unicompartmental osteoarthritis: a systematic studyANZ J SurgYear: 2007772142117388822
Hollinghurst D,Stoney J,Ward T,Gill HS,Newman JH,Murray DW,Beard DJ. No deterioration of kinematics and cruciate function 10 years after medial unicompartmental arthroplastyKneeYear: 200613440416973362
Kim SJ,MacDonald M,Hernandez J,Wixson RL. Computer assisted navigation in total knee arthroplasty: improved coronal alignmentJ ArthroplastyYear: 2005201233116214013
Konyves A,Willis-Owen C,Spriggins A. The long-term benfit of computer-assited surgical navigation in unicompartmental knee arthroplastyJ Orthop Surg ResYear: 201059421194426
Laskin RS. Unicompartmental knee replacement some unanswered questionsClin OrthopYear: 20013922677111716394
Lindstrand A,Stenstrom A,Ryd L,Toksvig-Larsen S. The introduction period of unicompartmental knee arthroplasty is critical: a clinical, clinical multicentered, and radiostereometric study of 251 Duracon unicompartmental knee arthroplastiesJ ArthroplastyYear: 2000156081610960000
Mason JB,Fehring TK,Estok R,Banel D,Fahrbach K. Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgeryJ ArthroplastyYear: 200722109710618078876
Moreland JR. Mechanisms of failure in total knee arthroplastyClin OrthopYear: 198822649643335107
Nicoll D,Rowley DI. Internal rotational error of the tibial component is a major cause of pain after total knee replacementJ Bone Joint Surg (Br)Year: 201092912384420798441
Perlick L,Bäthis H,Tingart M,Perlick C,Luring C,Grifka J. Minimally invasive unicompartmental knee replacement with a nonimage-based navigation systemInt Orthop (SICOT)Year: 2004281937
Repicci GD,Eberle RW. Minimally invasive surgery technique for unicondylar knee arthroplastyJ South Orthop AssocYear: 1999820712132848
Ridgeway SR,McAuley JP,Ammeen DJ,Engh GA. The effect of alignment of the knee on the outcome of unicompartmental knee replacementJ Bone Joint Surg (Br)Year: 2002843351512002491
Romanowski MR,Repicci JA. Minimally invasive unicondylar arthroplasty: eight-year follow-upJ Knee SurgYear: 200215172211829329
Rosenberger RE,Fink C,Quirbach S,Attal R,Tecklenburg K,Hoser C. The immediate effect of navigation on implant accuracy in primary mini-invasive unicompartmental knee arthroplastyKnee Surg Traumatol ArthroscYear: 20081612113340
Saccomanni B. Unicompartmental knee arthroplasty: a review of literatureClin RheumatolYear: 2010293394620069325
Schnurr C,Nessler J,Konig DP. Is referencing the posterior condyles sufficient to achieve a rectangular flexion gap in total knee arthroplasty?Int OrthopYear: 2009331561518956189
Servien E,Fary C,Lustig S,Demey G,Saffarini M,Chomel S,Neyret P. Tibial component rotation assessment using CT-scan in medial and lateral unicompartmental knee arthroplastyOrthop Tramatolog Surg ResYear: 20119732735
van der Linden-van der Zwaag HM,Bos J,van der Heide HJ,Nelissen RG. A computed tomography based study on rotational alignment accuracy of the femoral component in total knee arthroplasty using computer-assisted orthopaedic surgeryInt OrthopYear: 20113568455020623282
Weale AE,Murray DW,Baines J. Radiological changes 5 years efter unicompartmental knee replacementJ Bone Joint Surg (Br)Year: 20008299611041588

Figures

[Figure ID: F1]
Figure 1. 

Minimally invasive unicompartmental arthroplasty performed with computer-assisted surgery.



[Figure ID: F2]
Figure 2. 

Bony model created by the computer for positioning of the cutting block.



Tables
[TableWrap ID: T1] Table 1. 

Results for femoral component positioning in varus-valgus, extension-flexion, and rotation as measured with 3-D CT


Varus–
valgus
Extension–
flexion
External
rotation
Median CT value for the femoral implant position 11°
Positioning range for the femoral implant 4° to 4° 4° to 6° 6° to 20°
Average difference between CAS and CT a 1.2° 1.4° 9.9°
Outside of target range (≥ 3°) for the femoral implant b 3 1 8

T1-F1a The mean difference between the intraoperative computer-assisted measurement and the final

position postoperatively as calculated with CT.

T1-F2b Number of patients with an arbitrary difference of greater than 3 degrees between the desired

implant position intraoperatively and the postoperative CT measurement.


[TableWrap ID: T2] Table 2. 

Results for tibial component positioning in varus-valgus, flexion-extension, and rotation as measured with CT


Varus–
valgus
Extension–
flexion
External
rotation
Median CT values for the tibial implant position 10°
Positioning range for the tibial implant 7° to 2° 1° to 10° 4° to 28°
Average difference between CAS and CT a 2.2° 11°
Outside of target range (≥ 3°) for the tibial implant (2) b 1 2 6

T2-F1a The mean difference between the intraoperative computer-assisted measurement and the final

position postoperatively as calculated with CT.

T2-F2b Number of patients with an arbitrary difference of greater than 3 degrees between the desired

implant position intraoperatively and the postoperative CT measurement.



Article Categories: