Document Detail


Determining the adequate number of internal quality control levels: the example of coagulation factor VIII assay.
MedLine Citation:
PMID:  18600095     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Taking the specific case of coagulation factor VIII assay, we determined the characteristics of an internal quality control panel assuring control of the assay method for all of the critical factor VIII concentrations. The precision of the assay method was determined on six control materials C1-C6, with expected factor VIII levels of 1, 5, 30, 50, 80 and 150 U/dl, respectively. Given that, when two control levels correlate statistically, the information provided by one of them is redundant, we used correlation and principal components analysis to define a priori adequate and inadequate control panels. For each of these panels, we calculated the number of runs required, using Hotelling's method, to detect a shift expressed on C1 and impacted on C2, C3, C4, C5 and C6 in relation to the correlation phenomena among the six levels. The C1/C6 panel proved to be as informative in this regard as the complete panel for a 1 U/dl shift simulated on C1 and impacted on other levels too. These correlation phenomena allow the biologist to implement fewer control levels than there are critical concentrations needing to be explored in the internal quality control plan.
Authors:
Frédéric Sobas; Laurent Mazliak; Audrey Bellisario; Mathieu Lefranc; Anne Lienhart; Christophe Nougier; Claude Négrier
Related Documents :
9025135 - Basic framework for the economic evaluation of animal health control programmes.
21878405 - An atrioventricular node model for analysis of the ventricular response during atrial f...
16477985 - Current state and development of the real-time control of the berlin sewage system.
9452905 - A developmental transition in prehension modeled as a cusp catastrophe.
19906485 - On the minimum ignition energy (mie) for propane/air.
23742455 - Statistically testing the validity of analytical and computational approximations to th...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis     Volume:  19     ISSN:  0957-5235     ISO Abbreviation:  Blood Coagul. Fibrinolysis     Publication Date:  2008 Jul 
Date Detail:
Created Date:  2008-07-07     Completed Date:  2009-08-18     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9102551     Medline TA:  Blood Coagul Fibrinolysis     Country:  England    
Other Details:
Languages:  eng     Pagination:  433-7     Citation Subset:  IM    
Affiliation:
Haemostasis laboratory, Edouard Herriot Hospital, Lyon, France. frederic.sobas@chu-lyon.fr
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Blood Coagulation Tests / methods*,  standards*
Factor VIII / analysis*
Humans
Reference Standards
Chemical
Reg. No./Substance:
9001-27-8/Factor VIII