Detection of ventricular fibrillation in the presence of cardiopulmonary resuscitation artefacts.
MedLine Citation:
PMID:  17088016     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Providing cardiopulmonary resuscitation (CPR) to a patient in cardiac arrest introduces artefacts into the electrocardiogram (ECG), corrupting the diagnosis of the underlying heart rhythm. CPR must therefore be discontinued for reliable shock advice analysis by an automated external defibrillator (AED). Detection of ventricular fibrillation (VF) during CPR would enable CPR to continue during AED rhythm analysis, thereby increasing the likelihood of resuscitation success. This study presents a new adaptive filtering method to clean the ECG. The approach consists of a filter that adapts its characteristics to the spectral content of the signal exclusively using the surface ECG that commercial AEDs capture through standard patches. A set of 200 VF and 25 CPR artefact samples collected from real out-of-hospital interventions were used to test the method. The performance of a shock advice algorithm was evaluated before and after artefact removal. CPR artefacts were added to the ECG signals and four degrees of corruption were tested. Mean sensitivities of 97.83%, 98.27%, 98.32% and 98.02% were achieved, producing sensitivity increases of 28.44%, 49.75%, 59.10% and 64.25%, respectively, sufficient for ECG analysis during CPR. Although satisfactory and encouraging sensitivity values have been obtained, further clinical and experimental investigation is required in order to integrate this type of artefact suppressing algorithm in current AEDs.
Authors:
Elisabete Aramendi; Sofia Ruiz de Gauna; Unai Irusta; Jesus Ruiz; M Fe Arcocha; Jose Miguel Ormaetxe
Related Documents :
20228676 - Apoptosis is not involved in the mechanism of myocardial dysfunction after resuscitat...
6803436 - Ultrastructure of myocardium in the hurler syndrome. possible relation to cardiac funct...
8332576 - Close correlations between mitochondrial swelling and atp-content in the ischemic canin...
4053636 - Manual versus mechanical cardiopulmonary resuscitation in an experimental canine model.
17602036 - Milrinone enhances systolic, but not diastolic function during coronary artery bypass g...
10360716 - Ventricular radius reduction without resection: a computational analysis.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't     Date:  2006-11-07
Journal Detail:
Title:  Resuscitation     Volume:  72     ISSN:  0300-9572     ISO Abbreviation:  Resuscitation     Publication Date:  2007 Jan 
Date Detail:
Created Date:  2006-12-18     Completed Date:  2007-04-24     Revised Date:  2009-08-25    
Medline Journal Info:
Nlm Unique ID:  0332173     Medline TA:  Resuscitation     Country:  Ireland    
Other Details:
Languages:  eng     Pagination:  115-23     Citation Subset:  IM    
Affiliation:
Department of Electronics and Telecommunications, University of the Basque Country, Alameda de Urquijo s/n, 48013-Bilbao, Spain. jtparece@bi.ehu.es
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Algorithms
Artifacts
Electrocardiography / methods*
Humans
Sensitivity and Specificity
Ventricular Fibrillation / physiopathology*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Early life determinants of musculoskeletal sickness absence in a cohort of Norwegians born in 1967-1...
Next Document:  Absence of TREM2 polymorphisms in patients with Alzheimer's disease and Frontotemporal Lobar Degener...