Document Detail


Detailed electrical analysis of lead failures in a small-scaled right ventricular defibrillator lead: reality of Sprint Fidelis medical device recalls.
MedLine Citation:
PMID:  20515854     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
INTRODUCTION: This paper illustrates our experience with the Sprint Fidelis lead (SF, single coil model 6931). We investigated lead failure incidence, analysed for possible predictive factors and examined the efficacy of integrated early ICD warning systems.
METHODS: We analysed 181 consecutive patients with SF (follow up: 406 +/- 250 days). Left ventricular ejection fraction, age, gender, follow up, ICD indication, type of device, duration of implantation, and target vein used for implantation were evaluated as potential predictive factors of lead failure. Additionally, the predictive value of recommended impedance alert adaptations, the potential effects of the sensing integrity counter (SIC), and of the new lead integrity alert (LIA)(R) were studied.
RESULTS: Nine lead failures were identified. Lead failure occurred significantly more often in patients with single- and dual-chamber devices. None of the patients under cardiac resynchronisation therapy (CRT) had a lead failure (p=0.04). Seven failures (78%) became apparent through inappropriate shock interventions. Impedance alert adaptations did not prevent any inappropriate shock intervention, but the SIC and the activation of the LIA might have prevented inappropriate interventions. A fractured pace/sense ring conductor was identified as the most vulnerable part of the SF lead (in 7 failures, 78%).
CONCLUSION: We verified an increased failure rate in patients with the SF lead. Only patients with CRT devices were free from lead failure, suggesting a correlation with increased physical activity. The impedance alert reprogramming did not predict any inadequate shock interventions but LIA may become a new valuable tool for the early detection of lead failure signs.
Authors:
Jens Kreuz; Osman Balta; Sebastian Zimmer; Rolf Fimmers; Lars Lickfett; Fritz Mellert; Georg Nickenig; Joerg Otto Schwab
Related Documents :
5904984 - Occurrence of lymphopenia in heart failure.
20019604 - Heart failure in the ethnic minorities.
16837084 - Autoimmunity in myocardial infarction.
17517344 - Ve/vco2 slope in older heart failure patients with normal versus reduced ejection fract...
16638554 - Mechanical cardiac support in the young with the berlin heart excor pulsatile ventricul...
22808844 - Calcified left ventricular endomyocardial fibrosis.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Hellenic journal of cardiology : HJC = Hellēnikē kardiologikē epitheōrēsē     Volume:  51     ISSN:  2241-5955     ISO Abbreviation:  Hellenic J Cardiol     Publication Date:    2010 May-Jun
Date Detail:
Created Date:  2010-06-02     Completed Date:  2010-11-16     Revised Date:  2013-08-07    
Medline Journal Info:
Nlm Unique ID:  101257381     Medline TA:  Hellenic J Cardiol     Country:  Greece    
Other Details:
Languages:  eng     Pagination:  219-25     Citation Subset:  IM    
Affiliation:
Department of Medicine-Cardiology, University of Bonn, Sigmund-Freud-Strasse 25, Bonn, Germany. jens.kreuz@ukb.uni-bonn.de
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Defibrillators, Implantable*
Electricity
Equipment Design
Equipment Failure / statistics & numerical data*
Female
Humans
Male
Middle Aged
Safety-Based Medical Device Withdrawals*