Document Detail


Design and validation of the Grip-ball for measurement of hand grip strength.
MedLine Citation:
PMID:  22925584     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
The Grip-ball is a new dynamometer used to evaluate grip strength, as well as for use in home-based rehabilitation of the hand and forearm. The Grip-ball consists of pressure and temperature sensors and an electronic wireless communication system contained in an airtight ball. That can be inflated to different pressures. The device has advantages over standard dynamometers in that it looks like a simple ball, and can wirelessly communicate via Bluetooth to any compatible receiver, thus have potential to be used for clinical assessment and rehabilitation in a remote setting. The reliability and reproducibility of the device were assessed for the pressure sensor itself, as well as the relationship between the force applied and the pressure measured by the Grip-ball. The initial validation was performed using the pressure sensor without the ball in order to confirm the accuracy of the sensor used. A second validation study was conducted using the Grip-ball rather than just its sensor to examine the relationship between the pressure measured inside the ball and force applied. The results showed that there is a very good correlation (r=0.997, p<0.05) between the pressure measured by the Grip-ball sensor and that measured by a Vigorimeter, thus confirming the reliability of the sensor used in the Grip-ball. A quadratic regression equation was calculated in order to predict the force applied based on the pressure measured inside the ball, and the initial pressure to which the ball was inflated (R(2)=0.97, standard error 10.9N). Such a finding compares favourably with the variability inherent in Jamar recordings, thus indicating that the Grip-ball could be used to assess grip force. An industrial version of the Grip-ball, which is currently under development, will be able to be used for the entire range of grip force in the population.
Authors:
Rana Jaber; David J Hewson; Jacques Duchêne
Related Documents :
22268614 - Blood pressure variability as sign of autonomic imbalance in patients with idiopathic d...
21316304 - Tissue doppler em and instantaneous end-diastolic stiffness: validation against pressur...
22150944 - Pressure ulcers in older hospitalised patients and its impact on length of stay: a retr...
22033274 - Temporary blood pressure drop after bevacizumab administration is associated with clini...
24251274 - Association of sympathovagal imbalance with cardiovascular risks in overt hypothyroidism.
10762944 - Late clinical outcome after intracoronary palmaz-schatz stenting with high pressure bal...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2012-8-24
Journal Detail:
Title:  Medical engineering & physics     Volume:  -     ISSN:  1873-4030     ISO Abbreviation:  Med Eng Phys     Publication Date:  2012 Aug 
Date Detail:
Created Date:  2012-8-28     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9422753     Medline TA:  Med Eng Phys     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Copyright Information:
Copyright © 2012 IPEM. Published by Elsevier Ltd. All rights reserved.
Affiliation:
Institut Charles Delaunay, UMR CNRS 6279, Université de Technologie de Troyes, 12 rue Marie Curie, BP2060, 10010 Troyes, France.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Predicting atrial fibrillation inducibility in a canine model by multi-threshold spectra of the recu...
Next Document:  Anterior cruciate ligament (ACL) reconstruction with quadriceps tendon autograft and press-fit fixat...