Document Detail


Description of a flow optimized oxygenator with integrated pulsatile pump.
MedLine Citation:
PMID:  21092033     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
Extracorporeal membrane oxygenation (ECMO) is a well-established therapy for several lung and heart diseases in the field of neonatal and pediatric medicine (e.g., acute respiratory distress syndrome, congenital heart failure, cardiomyopathy). Current ECMO systems are typically composed of an oxygenator and a separate nonpulsatile blood pump. An oxygenator with an integrated pulsatile blood pump for small infant ECMO was developed, and this novel concept was tested regarding functionality and gas exchange rate. Pulsating silicone tubes (STs) were driven by air pressure and placed inside the cylindrical fiber bundle of an oxygenator to be used as a pump module. The findings of this study confirm that pumping blood with STs is a viable option for the future. The maximum gas exchange rate for oxygen is 48mL/min/L(blood) at a medium blood flow rate of about 300mL/min. Future design steps were identified to optimize the flow field through the fiber bundle to achieve a higher gas exchange rate. First, the packing density of the hollow-fiber bundle was lower than commercial oxygenators due to the manual manufacturing. By increasing this packing density, the gas exchange rate would increase accordingly. Second, distribution plates for a more uniform blood flow can be placed at the inlet and outlet of the oxygenator. Third, the hollow-fiber membranes can be individually placed to ensure equal distances between the surrounding hollow fibers.
Authors:
Ralf Borchardt; Peter Schlanstein; Jutta Arens; Roland Graefe; Fabian Schreiber; Thomas Schmitz-Rode; Ulrich Steinseifer
Related Documents :
19775263 - Design of a small centrifugal blood pump with magnetic bearings.
16156293 - Suction detection for the micromed debakey left ventricular assist device.
20097543 - The influence of expansion on intraoperative bone blood flow in multisegmental maxillar...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Artificial organs     Volume:  34     ISSN:  1525-1594     ISO Abbreviation:  Artif Organs     Publication Date:  2010 Nov 
Date Detail:
Created Date:  2010-11-24     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7802778     Medline TA:  Artif Organs     Country:  United States    
Other Details:
Languages:  eng     Pagination:  904-10     Citation Subset:  IM    
Copyright Information:
© 2010, Copyright the Authors. Artificial Organs © 2010, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.
Affiliation:
Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Helmholtz Institute, Aachen, Germany. borchardt@hia.rwth-aachen.de
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Extracorporeal membrane oxygenation following norwood stage 1 procedures at a single institution.
Next Document:  A newly developed miniaturized heart-lung machine--expression of inflammation in a small animal mode...