DOSIMAP: a high-resolution 2-D tissue equivalent dosemeter for linac QA and IMRT verification.
MedLine Citation:
PMID:  18757897     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
New generation of radiation therapy accelerators requires highly accurate dose measurements with high spatial resolution patterns. IMRT is especially demanding since the positioning accuracy of all the multi-leafs should be verified for each applied field and at any incidence. A new 2-D tissue equivalent dosemeter is presented with high spatial resolution that can fulfil these tasks. A plastic scintillator sheet is sandwiched between two polystyrene cubes, and the emitted light is observed by a high-resolution camera. A patented procedure allows efficient discrimination of the scintillation proportional to the dose from the parasitic Cerenkov radiation. This extraction made on the cumulated images taken during an irradiation field at a rate of 10 images s(-1) provides high-resolution mapping of the dose rate and cumulated dose in quasi real time. The dosemeter is tissue equivalent (ICRU-44) and works both for electrons and photons without complex parameter adjustment, since phantom and detector materials are identical. The calibration is simple and independent of the irradiation conditions (energy, fluence, quality and so on). The principle of the dosemeter and its calibration procedure are discussed in this paper. The results and, in particular, the dose depth profiles are compared with standard ionisation chamber measurements in polystyrene for both photons and electrons. Finally, the detector specifications are summarised and one example of complex IMRT field is discussed.
Authors:
V Collomb-Patton; P Boher; T Leroux; J-M Fontbonne; A Batalla; A Vela
Related Documents :
11143257 - New dosimetric approach for multidimensional dose evaluation in gamma knife radiosurger...
18782447 - The glaas algorithm for portal dosimetry and quality assurance of rapidarc, an intensit...
15770037 - Detection of urinary tract stones at low-radiation-dose ct with z-axis automatic tube c...
18633467 - Clinical evaluation of dose reduction on image quality of panoramic radiographs.
14615547 - Evidence for persistent flow and aqueous sedimentation on early mars.
18633727 - Relation between three-dimensional echocardiography derived left ventricular volume and...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't     Date:  2008-08-30
Journal Detail:
Title:  Radiation protection dosimetry     Volume:  131     ISSN:  0144-8420     ISO Abbreviation:  Radiat Prot Dosimetry     Publication Date:  2008  
Date Detail:
Created Date:  2008-11-04     Completed Date:  2009-04-16     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8109958     Medline TA:  Radiat Prot Dosimetry     Country:  England    
Other Details:
Languages:  eng     Pagination:  100-9     Citation Subset:  IM    
Affiliation:
ELDIM, 1185 Rue d'Epron, F14200 Hérouville Saint Clair, France.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Algorithms
Calibration
Electrons
Humans
Particle Accelerators*
Phantoms, Imaging
Photons
Quality Control
Radiometry / instrumentation*
Radiotherapy*
Radiotherapy Dosage*
Radiotherapy, Intensity-Modulated*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Optical radiation emissions from compact fluorescent lamps.
Next Document:  Evaluation of the calibration procedure of active personal dosemeters for interventional radiology.