Document Detail


Crosstalk triggered safety standby pacing associated with an improperly seated ventricular lead.
MedLine Citation:
PMID:  18025063     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Ventricular Safety Standby (VSS) was designed to prevent the potentially serious consequences of crosstalk in many dual-chamber pulse generators. In this report, we describe a patient in whom the VSS was activated in a special way. When the VSS presents during bipolar sensing configuration, the clinician should notice the proper electrical contact between leads and pulse generator.
Authors:
Qi Sun; Keping Chen; Shu Zhang
Related Documents :
16573403 - Cardiac arrhythmias: diagnosis and management. the bradycardias.
83633 - Partial pacemaker recycling of implanted qrs-inhibited pulse generators.
9489983 - Thrombolysis in myocardial infarction frame count in saphenous vein grafts.
Publication Detail:
Type:  Case Reports; Journal Article     Date:  2007-11-19
Journal Detail:
Title:  Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology     Volume:  10     ISSN:  1532-2092     ISO Abbreviation:  Europace     Publication Date:  2008 Jan 
Date Detail:
Created Date:  2008-01-17     Completed Date:  2008-03-06     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  100883649     Medline TA:  Europace     Country:  England    
Other Details:
Languages:  eng     Pagination:  75-6     Citation Subset:  IM    
Affiliation:
Chinese Academy of Medical Sciences and Peking Union Medical College, Centre of Arrhythmia Diagnosis and Treatment, Fuwai Hospital, Beijing 100037, China.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Cardiac Pacing, Artificial / adverse effects,  methods*
Equipment Failure
Equipment Safety
Female
Heart Atria / physiopathology
Heart Ventricles / physiopathology
Humans
Middle Aged
Pacemaker, Artificial / adverse effects*
Sick Sinus Syndrome / physiopathology*,  therapy*