Correction of conebeam CT values using a planning CT for derivation of the "dose of the day".
MedLine Citation:
PMID:  17936387     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND AND PURPOSE: Verification of the actually delivered 3D dose distribution during each treatment fraction ("dose of the day") is the most complete and clinical relevant "in-vivo" check of an IMRT treatment. To do this, during patient treatment portal dose images are routinely acquired with our electronic portal imaging device to derive the delivered fluence map for each treatment field. In addition, a conebeam CT scan is acquired just prior to treatment to derive the patient geometry at the time of treatment. However, the use of conebeam CT scans for dose calculation is hampered by inaccuracies in the conversion of CT values to electron densities due to an enlarged scatter contribution. MATERIALS AND METHODS: In this work, a method is described for mapping of Hounsfield Units of the planning CT to the conebeam CT scan, while accounting for non-rigidity in the anatomy, e.g. related to weight loss, in an approximate way. The method was validated for head and neck cancer patients by comparing dose distributions calculated using adjusted Hounsfield Units with a golden standard. RESULTS AND CONCLUSIONS: The observed dose differences were less than 1% in the majority of points, and in at least 96% of the points a 3D gamma analysis resulted in gamma values of less than 1 when applying a 2%/2mm criterion, showing that this straightforward approach allows for an accurate dose calculation based on conebeam CT scans.
Authors:
Mathilda van Zijtveld; Maarten Dirkx; Ben Heijmen
Related Documents :
17122577 - Extent and localization of changes in upper airway caliber with varying concentrations ...
17207697 - Ct colonography: automatic measurement of polyp diameter compared with manual assessmen...
21770987 - Computed tomography or magnetic resonance imaging-assisted partial hoof wall resection ...
16980687 - Kilovoltage ct using a linac-ct scanner combination.
25240997 - The use of high-speed, continuous, t2-weighted magnetic resonance sequences and saline ...
11230657 - Comparison between mr imaging and 99mtc mibi scintigraphy in the evaluation of recurren...
Publication Detail:
Type:  Comparative Study; Journal Article; Research Support, Non-U.S. Gov't     Date:  2007-10-23
Journal Detail:
Title:  Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology     Volume:  85     ISSN:  0167-8140     ISO Abbreviation:  Radiother Oncol     Publication Date:  2007 Nov 
Date Detail:
Created Date:  2007-12-06     Completed Date:  2008-04-01     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8407192     Medline TA:  Radiother Oncol     Country:  Ireland    
Other Details:
Languages:  eng     Pagination:  195-200     Citation Subset:  IM    
Affiliation:
Department of Radiation Oncology, Erasmus MC-Daniel den Hoed Cancer Center, Rotterdam, The Netherlands. M.vanzijtveld@erasmusmc.nl
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Cone-Beam Computed Tomography / methods*
Head and Neck Neoplasms / radiotherapy
Humans
Radiotherapy Dosage*
Radiotherapy Planning, Computer-Assisted*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Ethidium monoazide and propidium monoazide for elimination of unspecific DNA background in quantitat...
Next Document:  Tumour delineation and cumulative dose computation in radiotherapy based on deformable registration ...