Correcting for nonlinearity effects of continuous flow isotope ratio mass spectrometry across a wide dynamic range.
MedLine Citation:
PMID:  22279022     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
RATIONALE: Environmental and biological investigations may require samples that vary over a wide range of concentrations and isotope ratios, making measurements using continuous flow isotope ratio mass spectrometry (CF-IRMS) problematic due to nonlinear signal response. We therefore developed a mathematical approach for correcting nonlinearities over a wide range of sample concentrations and actual δ values.
METHODS: Dilution series for two standards were prepared in septum-capped vials and introduced into the mass spectrometer via the standard sampling pathway. Parameters for a nonlinear signal correction were determined by regression on measured isotope ratio vs. both signal strength and actual isotope ratio. We further extended the dynamic range by adjusting the position of an open split based on analyte concentration. Effects of the open split setting required additional mathematical correction.
RESULTS: The nonlinearities were corrected over a 100-fold range of sample concentrations and across a 600‰ change in isotope ratios (for δO(2) /N(2) values). The precision, measured as standard deviation, across the upper 90% of the concentration range was ±0.08‰, ±0.05‰, and ±2.6‰ for δ(18) O, δ(15) N, and δO(2) /N(2) values, respectively; the precision across the lower 10% of the range was ±0.22‰, ±0.07‰, and ±7.6‰, respectively. In all cases the linearity correction represented only a small fraction of these precision values.
CONCLUSIONS: The empirical correction described here provides a relatively simple yet effective way to increase the usable signal range for CF-IRMS applications. This improvement in dynamic range should be especially helpful for environmental and biological field studies, where sampling methods may be constrained by external factors. Copyright © 2012 John Wiley & Sons, Ltd.
Authors:
Ari Kornfeld; Travis W Horton; Dan Yakir; Matthew H Turnbull
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Rapid communications in mass spectrometry : RCM     Volume:  26     ISSN:  1097-0231     ISO Abbreviation:  Rapid Commun. Mass Spectrom.     Publication Date:  2012 Feb 
Date Detail:
Created Date:  2012-01-26     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8802365     Medline TA:  Rapid Commun Mass Spectrom     Country:  England    
Other Details:
Languages:  eng     Pagination:  460-8     Citation Subset:  IM    
Copyright Information:
Copyright © 2012 John Wiley & Sons, Ltd.
Affiliation:
School of Biological Sciences, University of Canterbury, Private Bag 4800, Christchurch 8140, New Zealand.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Application of the N(2) /Ar technique to measuring soil-atmosphere N(2) fluxes.
Next Document:  Production and fragmentation of negative ions from neutral N-linked carbohydrates ionized by matrix-...