Document Detail


Coronary stent implantation changes 3-D vessel geometry and 3-D shear stress distribution.
MedLine Citation:
PMID:  10899339     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Mechanisms of in-stent restenosis are not fully understood. Shear stress is known to play a role in plaque and thrombus formation and is sensitive to changes in regional vessel geometry. Hence, we evaluated the regional changes in 3-D geometry and shear stress induced by stent placement in coronary arteries of pigs.Methods. 3-D reconstruction was performed, applying a combined angiographic and IVUS technique (ANGUS), from seven Wallstents (diameter 3.5 (n=3) and 5mm (n=4)), which were implanted in seven coronary arteries of five pigs. This 3-D geometry was used to calculate locally the curvature, while the shear stress distribution was obtained by computational fluid dynamics. Local changes in shear stress were obtained at the entrance and exit of the stent for baseline (0. 65+/-0.22 ml/s) and hyperemic flow (2.60+/-0.86 ml/s) conditions. Results. After stent implantation, the curvature increased by 121% at the entrance and by 100% at the exit of the stent, resulting in local changes in shear stress. In general, at the entrance of the stent local maxima in shear stress were generated, while at the exit both local maxima and minima in shear stress were observed (p<0.05). Additionally, the shear stress at the entrance and exit of the stent were correlated with the local curvature (r: 0.30-0.84).Conclusion. Stent implantation changes 3-D vessel geometry in such a way that regions with decreased and increased shear stress occur close to the stent edges. These changes might be related to the asymmetric patterns of in-stent restenosis.
Authors:
J J Wentzel; D M Whelan; W J van der Giessen; H M van Beusekom; I Andhyiswara; P W Serruys; C J Slager; R Krams
Related Documents :
12036469 - The polytetrafluoroethylene-covered stent: a device with multiple potential advantages.
11146689 - Carotid artery stenting: utility of cerebral protection devices.
8951749 - Supra-aortic arterial stenoses: management with palmaz balloon-expandable intraluminal ...
19377499 - First case of stenting of a vulnerable plaque in the secritt i trial-the dawn of a new ...
18440199 - Successful repair of thoracoabdominal aortic aneurysm secondary to chronic dissection b...
15173879 - Getting stents to go with the flow.
12808679 - Arteriovenous malformation of the uterus associated with secondary postpartum hemorrhage.
9408629 - Aneurysm of the mitral-aortic intervalvular fibrosa as a rare cause of angina pectoris:...
22302579 - Superior mesenteric artery aneurysm in a 9-year-old boy with classical ehlers-danlos sy...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Journal of biomechanics     Volume:  33     ISSN:  0021-9290     ISO Abbreviation:  J Biomech     Publication Date:  2000 Oct 
Date Detail:
Created Date:  2000-10-04     Completed Date:  2000-10-04     Revised Date:  2009-11-11    
Medline Journal Info:
Nlm Unique ID:  0157375     Medline TA:  J Biomech     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  1287-95     Citation Subset:  IM    
Affiliation:
Thoraxcenter, University Hospital Dijkzigt, Erasmus University, Rotterdam, Netherlands.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Coronary Vessels / physiopathology*
Hemodynamics
Models, Cardiovascular*
Stents*
Stress, Mechanical
Swine

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Wear particle diffusion and tissue differentiation in TKA implant fibrous interfaces.
Next Document:  Surface roughness parameters as predictors of anchorage strength in bone: a critical analysis.