| Coronary sinus shocking lead as salvage in patients with advanced CHF and high defibrillation thresholds. | |
| | |
MedLine Citation:
|
PMID: 20230467 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
BACKGROUND: We report a series of three patients whose implantable cardioverter-defibrillators (ICD) implants were unsuccessful due to inability to achieve defibrillation thresholds (DFT) at maximum available energy after failure of standard modification and enhancement procedures. All patients had advanced cardiomyopathy. METHODS: Use of the coronary sinus (CS) for left ventricular (LV) shocking electrode placement resulted in acceptable DFTs in each patient. The position of the shocking coil in all three patients was posterior, and in two patients alongside a left ventricular CS pacing lead. The best shocking configuration tested was LV (CS) + CAN (Anode) to RV (cathode) in each patient. The short- and long-term outcomes of these patients is presented and discussed. CONCLUSION: This approach is suggested as a salvage option for those problematic patients who have unacceptable DFT results at implantation of an endovascular ICD system. |
| | |
Authors:
|
Osman Faheem; Arun Padala; Jeffrey Kluger; Steven Zweibel; Christopher A Clyne |
Related Documents
:
|
3045477 - Idiopathic ventricular tachycardia. a review. 3728287 - Laser ablation of normal and diseased human ventricle. 21094367 - Effect of implantable cardioverter-defibrillator on left ventricular ejection fraction ... 11446507 - Inappropriate discharges from an intravenous implantable cardioverter defibrillator due... 12827007 - Autoimmune thyroid disease and celiac disease in children. 8570877 - The impact of long-term hemodialysis on pituitary-adrenocortical function. |
Publication Detail:
|
Type: Case Reports; Journal Article Date: 2010-03-08 |
Journal Detail:
|
Title: Pacing and clinical electrophysiology : PACE Volume: 33 ISSN: 1540-8159 ISO Abbreviation: Pacing Clin Electrophysiol Publication Date: 2010 Aug |
Date Detail:
|
Created Date: 2010-08-13 Completed Date: 2010-12-06 Revised Date: - |
Medline Journal Info:
|
Nlm Unique ID: 7803944 Medline TA: Pacing Clin Electrophysiol Country: United States |
Other Details:
|
Languages: eng Pagination: 967-72 Citation Subset: IM |
Affiliation:
|
Division of Cardiology, Hartford Hospital, Hartford, Connecticut, USA. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Adult Aged Coronary Sinus* Defibrillators, Implantable Electric Countershock* / methods Electrodes, Implanted Heart Catheterization Heart Failure / physiopathology, therapy* Humans Male Middle Aged Salvage Therapy |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Catheter ablation of ventricular insertion of Mahaim fiber using noncontact mapping system in a pati...