Coronary CT angiography incorporating Doppler-guided prospective ECG gating in patients with high heart rate: comparison with results of traditional prospective ECG gating.
Jump to Full Text
MedLine Citation:
PMID:  23696793     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
PURPOSE: As Doppler ultrasound has been proven to be an effective tool to predict and compress the optimal pulsing windows, we evaluated the effective dose and diagnostic accuracy of coronary CT angiography (CTA) incorporating Doppler-guided prospective electrocardiograph (ECG) gating, which presets pulsing windows according to Doppler analysis, in patients with a heart rate >65 bpm.
MATERIALS AND METHODS: 119 patients with a heart rate >65 bpm who were scheduled for invasive coronary angiography were prospectively studied, and patients were randomly divided into traditional prospective (n = 61) and Doppler-guided prospective (n = 58) ECG gating groups. The exposure window of traditional prospective ECG gating was set at 30%-80% of the cardiac cycle. For the Doppler group, the length of diastasis was analyzed by Doppler. For lengths greater than 90 ms, the pulsing window was preset during diastole (during 60%-80%); otherwise, the optimal pulsing intervals were moved from diastole to systole (during 30%-50%).
RESULTS: The mean heart rates of the traditional ECG and the Doppler-guided group during CT scanning were 75.0±7.7 bpm (range, 66-96 bpm) and 76.5±5.4 bpm (range: 66-105 bpm), respectively. The results indicated that whereas the image quality showed no significant difference between the traditional and Doppler groups (P = 0.42), the radiation dose of the Doppler group was significantly lower than that of the traditional group (5.2±3.4 mSv vs. 9.3±4.5 mSv, P<0.001). The sensitivities of CTA applying traditional and Doppler-guided prospective ECG gating to diagnose stenosis on a segment level were 95.5% and 94.3%, respectively; specificities 98.0% and 97.1%, respectively; positive predictive values 90.7% and 88.2%, respectively; negative predictive values 99.0% and 98.7%, respectively. There was no statistical difference in concordance between the traditional and Doppler groups (P = 0.22).
CONCLUSION: Doppler-guided prospective ECG gating represents an improved method in patients with a high heart rate to reduce effective radiation doses, while maintaining high diagnostic accuracy.
Authors:
Min Li; Bing-bing Yu; Jian-hua Wu; Lin Xu; Gang Sun
Related Documents :
23803233 - Decision making and results of coronary artery bypass grafting for patients with poor l...
23009223 - Total cholesterol content of erythrocyte membranes is associated with the severity of c...
20089483 - Sudden cardiac death in young competitive athletes due to genetic cardiac abnormalities.
Publication Detail:
Type:  Comparative Study; Journal Article; Randomized Controlled Trial     Date:  2013-05-16
Journal Detail:
Title:  PloS one     Volume:  8     ISSN:  1932-6203     ISO Abbreviation:  PLoS ONE     Publication Date:  2013  
Date Detail:
Created Date:  2013-05-22     Completed Date:  2013-12-23     Revised Date:  2014-02-06    
Medline Journal Info:
Nlm Unique ID:  101285081     Medline TA:  PLoS One     Country:  United States    
Other Details:
Languages:  eng     Pagination:  e63096     Citation Subset:  IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Coronary Angiography / methods*
Electrocardiography / methods*
Female
Heart Rate / physiology*
Humans
Male
Middle Aged
Prospective Studies
Radiation Dosage
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): PLoS One
Journal ID (iso-abbrev): PLoS ONE
Journal ID (publisher-id): plos
Journal ID (pmc): plosone
ISSN: 1932-6203
Publisher: Public Library of Science, San Francisco, USA
Article Information
Download PDF
Copyright: 2013 Li et al
License:
Received Day: 28 Month: 12 Year: 2012
Accepted Day: 28 Month: 3 Year: 2013
collection publication date: Year: 2013
Electronic publication date: Day: 16 Month: 5 Year: 2013
Volume: 8 Issue: 5
E-location ID: e63096
PubMed Id: 23696793
ID: 3656032
Publisher Id: PONE-D-13-01199
DOI: 10.1371/journal.pone.0063096

Coronary CT Angiography Incorporating Doppler-Guided Prospective ECG Gating in Patients with High Heart Rate: Comparison with Results of Traditional Prospective ECG Gating Alternate Title:CTA with Doppler-Guided Prospective ECG Gating
Min Li1
Bing-bing Yu1
Jian-hua Wu2
Lin Xu3
Gang Sun1*
Matthias Stuberedit1 Role: Editor
1Department of Medical Imaging, Jinan Military General Hospital, Jinan, Shandong Province, China
2School of Bioscience and Bioengineering, South China University of Technology, Guangzhou, Guangdong Province, China
3Department of Medical Cardiology, Jinan Military General Hospital, Shandong Province, China
Centre Hospitalier Universitaire Vaudois, Switzerland
Correspondence: * E-mail: cjr.sungang@vip.163.com
[conflict] Competing Interests: The authors have declared that no competing interests exist.
Contributed by footnote: Conceived and designed the experiments: GS. Performed the experiments: ML BY. Analyzed the data: JW. Contributed reagents/materials/analysis tools: ML BY LX. Wrote the paper: ML BY JW LX GS.

Introduction

With the clinical application of multi-detector CT scanning for imaging the coronary arteries, coronary CT angiography (CTA) has emerged as an attractive noninvasive diagnostic modality for detecting coronary artery disease [1]. However, since Einstein et al. estimated a significant number of potential radiation-induced neoplasms from coronary CTA [2], the high effective dose and potential adverse consequences of coronary CTA have aroused greater attention and have limited the general application of the technique [3].

Until now, many methods have been developed and applied to reduce the radiation dose, among which prospective electrocardiograph (ECG) gating is one effective method that has demonstrated the obvious advantage of allowing for a decreased patient dose [4][8]. However, it is only effective when the heart rate is low [9]. Traditionally, when the heart rate <65 bpm, diastole is recommended as the optimal pulsing window for a relatively longer phase with less motion of the coronary arteries. As the heart rate increases, however, the lengths of R-R intervals are shortened correspondingly, with diastole compressed significantly. As a result, it is difficult to produce an optimal image quality exclusively at diastole, as some patients should be scanned during systole [10][15]. In this case, the pulsing windows are usually expanded to cover both systole and diastole, which results in a higher radiation dose (1, 17, 18).

Numerous studies have mainly focused on the relationship between the heart rate and the optimal pulsing windows to guide the clinical utility of prospective ECG gating for patients with a high heart rate. The results, nevertheless, are contradictory, as the optimal pulsing window does not always change regularly with the heart rate. For example, Leschka et al. believed that the narrowest reconstruction window providing diagnostic image quality was 60–70% for a heart rate <60 bpm, 60–80% for a heart rate of 60–70 bpm, 55–80% for a heart rate of 70–80 bpm, and 30–80% for a heart rate >80 bpm (20), whereas Weustink et al. concluded the optimal exposure windows to be at 60–76%, 30–77%, and 31–47% for a heart rate ≤65 bpm, 66–79 bpm, and ≥80 bpm, respectively (19, 29). The possible reason may be that the physiological phase is associated not only with heart rate but also with other factors, such as filling pressures and heart function [16]. A β-blocker is therefore recommended to control the heart rate. Although using a β-blocker has been proven to be an effective approach with which to lower the patient's heart rate, it is ineffective and may have adverse effects in some patients [1].

A precise definition of an optimal ECG pulsing window is therefore required for prospective ECG gating scanning in patients with a high heart rate. Since Doppler ultrasound has been proven to be a helpful methodology with which to predict the physiological phase [17], it is hypothesized that, compared with the traditional prospective ECG-gating protocol, prospective ECG gating incorporating Doppler ultrasound analysis (Doppler-guided prospective ECG gating) has the potential to compress pulsing windows from both systole and diastole to systole or diastole in order to reduce patient dose [17]. In the present study, we evaluated the patient dose and diagnostic accuracy of CTA incorporating Doppler-guided prospective ECG gating in patients whose heart rate could not be brought under 65 bpm.


Materials and Methods
Study Population

Our study protocol and radiation dose measurements were approved by the Ethics Committee of Jinan Military General Hospital (Protocol S1). Written informed consents, including information about the risk of radiation and iodine allergic reactions, were obtained from all of the participants.

In total, 1431 symptomatic patients who were scheduled for invasive coronary angiography (ICA) were prospectively studied. For patients with a heart rate >65 bpm, a β-blocker was traditionally applied in order to reduce the heart rate. Excluded from the study were 744 patients with a heart rate <65 bpm and 547 patients with a heart rate >65 bpm, whose heart rate could be brought under 65 bpm by the β-blocker. Among the 102 patients whose heart rate could not be brought under 65 bpm by the β-blocker as well as 38 patients with a heart rate >65 bpm and contraindications for β-blockers (asthma and severe chronic obstructive emphysema), the subjects were randomly divided into a traditional prospective ECG-gating group and a Doppler-guided prospective ECG-gating group. Patients who met the following criteria were excluded: refused to provide consent or withdrew for personal reasons (n = 2); demonstrated a heart rate variability of more than 20 bpm (n = 13); had undergone stent-graft and bypass surgery (n = 5); or were allergic to iodinated contrast agents (n = 1). Our final study cohort consisted of 119 patients, with 61 patients in the traditional prospective ECG-gating group and 58 in the Doppler-guided prospective ECG-gating group, respectively (Figure 1). No significant difference was found between the baseline characteristics of the two groups (traditional vs. Doppler) (Table 1).

CT Angiography Protocol

All CT examinations were performed using 320-detector CT (Aquilion ONE, Toshiba, Nasu, Japan) with a detector collimation of 320×0.5 mm. For CT coronary imaging, a 40–60 ml bolus of Iohexol (Omnipaque 350 mg/ml; Amersham-GE Healthcare, Shanghai, China) was injected into an antecubital vein through an 18-gauge catheter at an injection rate of 4–6 ml/sec, followed by 50 ml of saline solution.

CT scanning parameters were set according to body mass index (BMI): 100 kV and 350 mA for BMI <18; 100 kV and 400 mA for BMI between 19–24; and 120 kV and 450–500 mA for BMI >24. For heart rates between 65–79 bpm, the two-heart-beat acquisition mode was used. For heart rates ≥80 bpm, the three-heart-beat acquisition mode was applied in order to increase the effective temporal resolution and obtain adequate imaging quality. Using the multi-heart-beat acquisition mode, the effective temporal resolution was able to reach 87.5–58.3 ms.

The exposure window of traditional prospective ECG gating was adjusted at 30%–80% of the cardiac cycle according to previous studies of the optimal pulsing windows of CTA_ENREF_19 (Figure 2A) [14], [15]. The pulsing intervals of Doppler-guided prospective ECG gating were determined by Doppler ultrasound analysis 5–10 minutes before CT angiography (Figure 2B). Transmitral pulsed-Doppler flow data were recorded from the transthoracic apical four-chamber using a clinical echocardiographic imaging system (Vivid 7; GE, USA), which was equipped within the same room with the CT. The patients were examined on the examining table of the CT scanner. The ECG data and Doppler data were recorded synchronously. The length of diastasis in diastole with less motion velocity was evaluated integrating the ECG signal. The end of the E-wave corresponded to the onset of diastasis (a), and the beginning of the late diastolic filling peak velocity (A-wave) corresponded to the end of diastasis (b) [18]. All data were recorded using the absolute timing (ms) from the previous R-peak. The length of diastasis was calculated by (b-a). When the length of diastasis was more than 90 ms, the pulsing windows were manually preset during diastole (during 60%–80%); if it was less than 90 ms, at which point the length was too short to reconstruct an image with good quality, the optimal pulsing intervals were moved from diastole to systole (during 30%–50%) [17].

Reconstruction and Post Processing

Image reconstruction was performed with 0.5-mm slice thickness and 0.25-mm overlap. At the level of the middle right coronary artery (RCA), we used an absolute timing approach and reconstructed transverse images with a 10-ms step from the peak R-wave. When the interval with less motion artifact was determined, the whole-heart data were reconstructed. All reconstructed images were transferred to an independent workstation (Vitrea II FX, Vital Images, Minnetonka, MN, USA).

Image Analysis

Coronary segments were defined by a 15-segment model according to American Heart Association guidelines [19]. All reconstructed images were evaluated and classified by two independent radiologists, who were blinded to the randomized protocol and ICA results. Image quality was classified into 4 grades: Grade 1—no artifacts and clear delineation of the segment; Grade 2—minor artifacts and mild blurring of the segment; Grade 3—moderate artifacts and moderate blurring; and Grade 4—severe artifacts and segment too poor for evaluation. The degree of involved lumen stenosis was measured using the narrowest dimension of the lumen at the level of stenosis compared with the normal lumen diameter distally. The measurements were made with electronic calipers in the Vitrea II FX workstation. The extent of vessel stenosis was classified into atherosclerosis with stenosis less than 50% and with stenosis greater than 50%. Consensus agreement was used for any disagreements.

Invasive Coronary Angiography

Cardiac angiograms were performed using the conventional Judkin technique [20] within two weeks after CTA. Four views of the left coronary artery (LCA) and two views of the RCA were analyzed in consensus by two cardiologists. They were blinded to the CT results during analysis. Quantitative assessment of stenosis severity on angiograms was performed with the same criteria as those used for the CT data on the GE ADW4.2 workstation.

Evaluation of Radiation Dose of CTA and ICA

The dose length product (DLP) displayed on the dose report of the CT scanner was recorded. An effective dose was obtained using the equation: E = k×DLP (k = 0.029 mSv×mGy−1×cm−1, which was calculated specifically for the 320-detector CT) [21]. For ICA, the effective dose was estimated as a product of the dose-area product (DAP) of the diagnostic coronary scenes × a conversion factor (k = 0.22 mSv/mGy×cm2) [22].

Statistical Analysis

Statistical analysis was performed with the SPSS, version 16 software package for Windows (SPSS, Chicago, II, USA). A P-value of less than 0.05 was considered statistically significant.

Quantitative variables were expressed as mean ± SD, and categorical variables as frequencies or percentages. An inter-observer agreement for the determination of image quality was calculated with kappa statistics.

Results of ICA were used as the reference standard to calculate the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of CTA. A McNemar test was conducted to evaluate the difference between CTA and ICA results. The Fisher exact test and chi-square test, as appropriate, were used to evaluate the concordances for the groups based on the two types of prospective ECG gating (traditional vs. Doppler). In a sub-analysis, patients were subdivided into two groups according to mean heart rate (group A: 65 ≤ heart rate <80 bpm; group B: heart rate ≥80 bpm). Three-way ANOVA was performed to evaluate the radiation exposure, with groups (traditional vs. Doppler), heart rate (65 ≤ heart rate <80 bpm vs. heart rate ≥80 bpm), and imaging methodology (CTA vs. ICA) as factors.


Results

The mean heart rate of the traditional-ECG gating group during CT scanning was 75.0±7.7 bpm (range, 66–96 bpm), with 72.4% (42/58) of a heart rate of 65 ≤ heart rate <80 bpm and 27.6% (16/58) of a heart rate ≥80 bpm. Results revealed a similar scanning heart rate for the Doppler ECG-gating group (76.5±5.4 bpm; range: 66–105 bpm, P = 0.32), with 80.3% (49/61) and 19.7% (12/61) in heart rates of the group for which 65 ≤ heart rate <80 bpm group and the groups for which heart rate ≥80 bpm, respectively. In the Doppler group, 15 patients were scanned during systole (30%–50%), and 43 patients were exposed during diastole (60%–80%).

Image Quality

Of the 1,785 coronary arterial segments in 119 patients (15 segments per patient), 53 segments were excluded because the segments did not exist and could not be visualized by ICA and CTA. A total of 1,732 segments were thus available for analysis. The inter-observer agreement for image quality assessment was high (kappa = 0.78). The image quality grades of these segments were as follows: grade 1—81.9% (1419/1732); grade 2—12.9% (224/1732); grade 3—4.3% (74/1732); and grade 4—0.9% (9/1732). There was no significant difference in image quality score between traditional and Doppler-guided prospective ECG gating (P = 0.42) (Table 2).

Effective Dose

Overall, the effective dose in the Doppler-guided prospective ECG-gating group was significantly lower than that in the traditional prospective ECG-gating group (F = 4.6, P = .003), and the patient dose in ICA was significantly higher than that of CTA (F = 123.1, P<0.001) (Figure 3). In addition, the radiation dose in patients with a heart rate ≥80 bpm was higher than that in patients with a heart rate between 65 bpm and 80 bpm (F = 15.7, P<0.001).

A groups × imaging methodology interaction followed by separate ANOVAs for traditional and Doppler-guided prospective ECG gating revealed that, whereas the effective dose of ICA was similar for the two groups (15.3±3.6 mSv vs. 15.0±5.2 mSv, respectively, P = 0.65), the radiation dose of CTA in the Doppler-guided prospective ECG-gating group (5.2±3.4 mSv) was lower than that of traditional prospective ECG gating (9.3±4.5 mSv, P = 0.001) (Figure 3). This interaction was further qualified by a groups × imaging methodology × heart rate interaction, which suggested that, whereas the dose of CTA in traditional group was lower than that of ICA at heart rates between 65 to 80 bpm (8.2±5.3mSv vs. 14.9±5.6mSv, P<0.001), the dose of CTA increased significantly in heart rates ≥80 bpm, with no significant difference from that of ICA (13.5±4.1 mSv vs. 15.1±3.5mSv, P = 0.1). In the Doppler group, on the other hand, the dose of CTA was lower than that of ICA in both the group with a heart rate of 65 ≤ heart rate <80 bpm (4.6±3.3mSv vs. 14.5±3.2mSv, P<0.001) and the group with a heart rate ≥80 bpm (6.9±3.2mSv vs. 17.4±4.1mSv, P<0.001).

Diagnostic Accuracy

The inter-rater concordance between the two radiologists for ICA was good (kappa = 0.82). In the traditional prospective group, there were 154 segments in 44 patients with stenosis ≥50%. Among the 154 stenotic segments proven by ICA, CTA diagnosed 147 cases accurately. Among the 733 unstenotic segments proven by ICA, CTA diagnosed 718 cases accurately. The sensitivity of CTA was 95.5% (95% CI: 90.9, 97.8); the specificity was 98.0% (95% CI: 96.7, 98.8); the PPV was 90.7% (95% CI: 85.3, 94.3); and the NPV was 99.0% (95% CI: 98.0, 99.5), respectively. The McNemar test showed no significant difference in the diagnostic results between CTA and ICA (P = 0.13).

In the Doppler-guided prospective ECG-gating group, 158 segments in 42 patients were diagnosed with stenosis ≥50% by ICA, among which 149 segments were detected by CTA. In the 687 segments with stenosis <50%, 667 segments were diagnosed accurately by CTA. For the 845 evaluated segments, the sensitivity of CTA was 94.3% (95% CI: 89.5, 97.0); the specificity was 97.1% (95% CI: 95.6, 98.2); the PPV was 88.2% (95% CI: 82.4, 92.2); and the NPV was 98.7% (95% CI: 97.5, 99.3), respectively. The McNemar test showed no significant difference in the diagnostic results between CTA and ICA (P = 0.06) (Figure 4). The diagnostic accuracies at patient and artery levels are presented in Table 3. In total, there was no statistical difference in concordance between the traditional and Doppler-guided prospective ECG-gating groups at patient, artery, and segment levels (P = 1.0, 0.48, and 0.22, respectively).


Discussion

Although prospective ECG gating has been suggested as a useful procedure to reduce radiation dose, it only works when the heart rate is low. A β-blocker is therefore recommended before CTA to control the heart rate, but is ineffective in some patients [9]. In this study, we aimed to evaluate the clinical performance of CTA incorporating Doppler-guided prospective ECG gating in patients with a high heart rate. The results indicated that compared with traditional prospective ECG gating, prospective ECG gating incorporating Doppler analysis could decrease radiation dose significantly while maintaining high diagnostic accuracy, especially for patients in whom the β-blocker proves ineffective and might have adverse effects.

Due to the consistent movement of the heart, it is critical to find a relatively tranquil duration in which to scan and reconstruct the coronary artery. Studies have confirmed that there are two durations suitable for image reconstruction. One corresponds to the diastasis in diastole, while the other takes place in systole [23]. Theoretically, when the heart rate ≤65 bpm, it is better to assess vessel lumen in diastole due to the increased blood flow, vasodilatation effect, and comparatively longer duration [13]. However, when the heart rate is increasing, the optimal pulsing windows fluctuate, rather than locating at diastole or systole stably or changing regularly with heart rate [10][12]. Thus, at a high heart rate, the optimal pulsing windows should cover both systole and diastole in order to maintain image quality, resulting in a significantly increased patient dose. Incorporating the results of previous studies, we selected 30%–80% as the pulsing windows for traditional prospective ECG gating [15].

The present study showed that Doppler-guided prospective ECG gating, which can analyze the true length of physiological phase, may offer the possibility of locating pulsing windows exclusively at systole or diastole and thus decrease radiation dose. Previous study has demonstrated that when the length of diastasis >90 ms, the optimal image quality could be obtained at diastole. Otherwise, when the length of diastasis <90 ms, the diastole is not suitable for optimal reconstruction, and the reconstruction windows should be set at systole [17]. As the pulsing windows were compressed from 30%–80% to 60%–80% or 30%–50%, the Doppler-guided prospective ECG gating could reduce radiation dose by nearly 44% of patient dose.

It is worth noting that in our study, the effective dose of Doppler-guided prospective ECG gating was still higher than those of previous studies based on patients with a low heart rate (5.2mSv vs. 2.2–4.2mSv) [1], [24], [25]. However, the present study showed a significant lower dose in patients with a high heart rate. In a study using 320-detector CT, when the heart rate >65 bpm, the radiation dose increased from 3.9 mSv in patients with a heart rate <65 bpm to 12.3 mSv, which was also two times more than the dose for Doppler-guided prospective ECG gating [1].

Generally, three main factors influence the radiation dose in patients with a high heart rate. The first is the length of exposure window. As mentioned above, this new technique of Doppler-guided prospective ECG gating cardiac CT angiography has the ability to compass the pulsing windows from 30%–80% to 60%–80% or 30%–50%. The second is the temporal resolution, determined by the hardware of CT scanners. If the temporal resolution is not sufficient to reconstruct a diagnostic image in patients with a high heart rate, the multi-segment reconstruction method should be applied to improve the effective temporal resolution, which increases the exposure time and patient dose significantly. Although the exposure windows were compressed greatly by Doppler-guided prospective ECG gating in the present study, the multi-segment reconstruction was applied by 320-detector CT, which led to a redundant dose when the heart rate is high [1]. As we can see from the results, the dose in patients with a heart rate >80 bpm was significantly higher than that in patients with a heart rate between 65 and 80 bpm for the application of the three-heart-beat acquisition mode. It is possible that dual-source CT (DSCT), with an advanced temporal resolution of 82.5 ms, could reduce the radiation dose further by applying Doppler-guided prospective ECG gating [26]. The third factor is the application of the iterative algorithm. The iterative reconstruction algorithm allows the dose to be lowered while maintaining image quality [27]. As the iterative algorithm was not applied in the first generation of 320-detectors CT, the radiation dose was still comparatively high. A recent study on the second-generation 320-detector CT scanner with iterative reconstruction in patients with a low heart rate showed a significantly reduced radiation dose, which could decrease radiation dose still further than the present CT scanner [28]. Therefore, the reduced radiation dose should not depend on one technique alone, but on a synthetic application of low dose techniques. Further study combining narrower pulsing windows, higher temporal resolution, and an iterative reconstruction algorithm may decrease the requisite radiation dose further.

Another reason for the redundant dose of the study was the differing conversion factor in dose estimation. Previous studies usually applied k = 0.014 mSv×mGy−1×cm−1 or k = 0.017 mSv×mGy−1×cm−1 as the conversion coefficient, which however is chest conversion factor. Recent studies have suggested that the conversion coefficient changes depending on organ, scanning mode, patient size, X-ray tube voltage, and other factors [29]. Applying dual-source CT, Hurwitz et al. suggested that the conversion factor for cardiac CT should be k = 0.025 mSv×mGy−1×cm−1[30]. As mentioned above, Einstein et al. suggested that the conversion factor (k) should be 0.029 mSv×mGy−1×cm−1 for 320-detectors CT [21], which is significantly higher than the conversion factor used in previous studies.

It is worth noting that the procedure of CTA with Doppler-guided prospective ECG gating was more time-consuming and costly than the traditional method. However, patients whose heart rate could not be brought to desired levels only made up 8.3% of the total patients and thus did not disturb the work routine significantly. With regard to individual protection from radioactivity, Doppler-guided prospective ECG gating may reduce the patient dose by nearly 44%. Further studies should be conducted to investigate the cost-effectiveness of Doppler-guided prospective ECG gating in order to prove that the 44% of radiation dose saved benefits long-term survival.

There are two limitations to our study. First, patients with arrhythmia were excluded from this study, as it was difficult for Doppler to forecast the tranquil duration interval. Second, the study was performed by applying first-generation 320-detector CT. In the future, the advantages of Doppler-guided prospective ECG gating should be confirmed with other scanners.

In conclusion, in terms of decreasing the patient dose, ECG gating for coronary CTA improves from retrospective without dose modulation to retrospective with dose modulation to prospective mode. The current study indicates that Doppler-guided prospective ECG gating, a new method of ECG gating, may reduce effective doses further, while maintaining high diagnostic accuracy, particularly in patients with an elevated heart rate.


Supporting Information Protocol S1

The Detailed Protocol Approved by the Ethics Committee of Jinan Military General Hospital.

(DOC)


Click here for additional data file (pone.0063096.s001.doc)


References
1. Dewey M,, Zimmermann E,, Deissenrieder F,, Laule M,, Dubel HP,, et al. (Year: 2009) Noninvasive coronary angiography by 320-row computed tomography with lower radiation exposure and maintained diagnostic accuracy: comparison of results with cardiac catheterization in a head-to-head pilot investigation. Circulation120: 867–87519704093
2. Einstein A,, Henzlova M,, Rajagopalan S, (Year: 2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. Jama298: 31717635892
3. Redberg R,, Walsh J, (Year: 2008) Pay now, benefits may follow–the case of cardiac computed tomographic angiography. N Engl J Med359: 2309–231119038877
4. Paul JF,, Abada HT, (Year: 2007) Strategies for reduction of radiation dose in cardiac multislice CT. Eur Radiol17: 2028–203717318604
5. Husmann L,, Valenta I,, Gaemperli O,, Adda O,, Treyer V,, et al. (Year: 2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J29: 191–19718089704
6. Horiguchi J,, Kiguchi M,, Fujioka C,, Shen Y,, Arie R,, et al. (Year: 2008) Radiation dose, image quality, stenosis measurement, and CT densitometry using ECG-triggered coronary 64-MDCT angiography: a phantom study. AJR Am J Roentgenol190: 31518212215
7. Earls J,, Berman E,, Urban B,, Curry C,, Lane J,, et al. (Year: 2008) Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: improved image quality and reduced radiation dose. Radiology246: 742–75318195386
8. Hirai N,, Horiguchi J,, Fujioka C,, Kiguchi M,, Yamamoto H,, et al. (Year: 2008) Prospective versus retrospective ECG-gated 64-Detector coronary CT angiography: assessment of image quality, stenosis, and radiation dose. Radiology248: 424–43018574140
9. von Ballmoos MW,, Haring B,, Juillerat P,, Alkadhi H, (Year: 2011) Meta-analysis: diagnostic performance of low-radiation-dose coronary computed tomography angiography. Ann Intern Med154: 413–42021403076
10. Seifarth H,, Wienbeck S,, Pusken M,, Juergens KU,, Maintz D,, et al. (Year: 2007) Optimal systolic and diastolic reconstruction windows for coronary CT angiography using dual-source CT. AJR Am J Roentgenol189: 1317–132318029865
11. Leschka S,, Husmann L,, Desbiolles LM,, Gaemperli O,, Schepis T,, et al. (Year: 2006) Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT. Eur Radiol16: 1964–197216699752
12. Leschka S,, Wildermuth S,, Boehm T,, Desbiolles L,, Husmann L,, et al. (Year: 2006) Noninvasive coronary angiography with 64-section CT: effect of average heart rate and heart rate variability on image quality. Radiology241: 378–38516966483
13. Steigner ML,, Otero HJ,, Cai T,, Mitsouras D,, Nallamshetty L,, et al. (Year: 2009) Narrowing the phase window width in prospectively ECG-gated single heart beat 320-detector row coronary CT angiography. Int J Cardiovasc Imaging25: 85–9018663599
14. Weustink AC,, Mollet NR,, Pugliese F,, Meijboom WB,, Nieman K,, et al. (Year: 2008) Optimal electrocardiographic pulsing windows and heart rate: effect on image quality and radiation exposure at dual-source coronary CT angiography. Radiology248: 792–79818710975
15. Leschka S,, Scheffel H,, Desbiolles L,, Plass A,, Gaemperli O,, et al. (Year: 2007) Image quality and reconstruction intervals of dual-source CT coronary angiography: recommendations for ECG-pulsing windowing. Invest Radiol42: 543–54917620936
16. Van Kraaij D,, Van Pol P,, Ruiters A,, De Swart J,, Lips D,, et al. (Year: 2002) Diagnosing diastolic heart failure. Eur J Heart Fail Failure4: 419–430
17. Sun G,, Li M,, Jiang XS,, Li L,, Peng ZH,, et al. (Year: 2011) Transthoracic Doppler echocardiography to predict optimal tube pulsing window for coronary artery CT angiography. Eur J Radiol81: 2212–2220
18. Chung C,, Karamanoglu M,, Kovacs S, (Year: 2004) Duration of diastole and its phases as a function of heart rate during supine bicycle exercise. Am J Physiol Heart Circ Physiol287: H2003–200815217800
19. Knez A,, Becker CR,, Leber A,, Ohnesorge B,, Becker A,, et al. (Year: 2001) Usefulness of multislice spiral computed tomography angiography for determination of coronary artery stenoses. Am J Cardiol88: 1191–119411703970
20. Shah A,, Gnoj J,, Fisher V, (Year: 1975) Complications of selective coronary arteriography by the Judkins technique and their prevention. Am Heart J90: 353–3591163427
21. Einstein AJ,, Elliston CD,, Arai AE,, Chen MY,, Mather R,, et al. (Year: 2010) Radiation dose from single-heartbeat coronary CT angiography performed with a 320-detector row volume scanner. Radiology254: 698–70620177085
22. Herzog BA,, Wyss CA,, Husmann L,, Gaemperli O,, Valenta I,, et al. (Year: 2009) First head-to-head comparison of effective radiation dose from low-dose 64-slice CT with prospective ECG-triggering versus invasive coronary angiography. Heart95: 1656–166119581273
23. Herzog C,, Arning-Erb M,, Zangos S,, Eichler K,, Hammerstingl R,, et al. (Year: 2006) Multi-detector row CT coronary angiography: influence of reconstruction technique and heart rate on image quality. Radiology238: 75–8616373760
24. Shuman WP,, Branch KR,, May JM,, Mitsumori LM,, Lockhart DW,, et al. (Year: 2008) Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries: comparison of image quality and patient radiation dose. Radiology248: 431–43718552312
25. Stolzmann P,, Goetti R,, Baumueller S,, Plass A,, Falk V,, et al. (Year: 2010) Prospective and retrospective ECG-gating for CT coronary angiography perform similarly accurate at low heart rates. Eur J Radiol79: 85–9120079993
26. Lell M,, Marwan M,, Schepis T,, Pflederer T,, Anders K,, et al. (Year: 2009) Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT: technique and initial experience. Eur Radiol19: 2576–258319760421
27. Beister M,, Kolditz D,, Kalender WA, (Year: 2012) Iterative reconstruction methods in X-ray CT. Phys Med28: 94–10822316498
28. Chen MY,, Shanbhag SM,, Arai AE, (Year: 2013) Submillisievert median radiation dose for coronary angiography with a second-generation 320-detector row CT scanner in 107 consecutive patients. Radiology267: 76–8523340461
29. Huda W,, Yoshizumi T,, Hurwitz L,, Goodman P, (Year: 2008) Computing effective doses from dose-length product in CT. Radiology248: 32118566186
30. Hurwitz LM,, Reiman RE,, Yoshizumi TT,, Goodman PC,, Toncheva G,, et al. (Year: 2007) Radiation dose from contemporary cardiothoracic multidetector CT protocols with an anthropomorphic female phantom: implications for cancer induction. Radiology245: 742–75017923509

Figures

[Figure ID: pone-0063096-g001]
doi: 10.1371/journal.pone.0063096.g001.
Figure 1  Flow Chart of References Searching.

The diagram shows the exact criteria for the inclusion and exclusion of subjects.



[Figure ID: pone-0063096-g002]
doi: 10.1371/journal.pone.0063096.g002.
Figure 2  Methods of Traditional and Doppler-Guided Prospective ECG Gating.

A) presents a schematic view of traditional prospective ECG gating. The exposure window is adjusted to cover both systole and diastole. B) offers a schematic view of Doppler-guided prospective ECG gating. The exposure window was adjusted at systole or diastole.



[Figure ID: pone-0063096-g003]
doi: 10.1371/journal.pone.0063096.g003.
Figure 3  Dose of Traditional and Doppler-Guided Prospective ECG Gating.

Overall, applying Doppler-guided prospective ECG gating, the patient dose of CTA was significantly lower than that of traditional prospective ECG gating.



[Figure ID: pone-0063096-g004]
doi: 10.1371/journal.pone.0063096.g004.
Figure 4  Example of CTA with Doppler-Guided Prospective ECG Gating.

A male patient with a BMI of 21.5. The predicted length of DTD was 81 ms, and the exposure timing was preset at 30%–50% during the R-R interval. The scanning heart rate was 78 bpm. CTA with Doppler-guided prospective ECG gating (A) and ICA (B) showed stenosis of LAD (white arrows). The effective dose of CTA was 3.2 mSv.



Tables
[TableWrap ID: pone-0063096-t001] doi: 10.1371/journal.pone.0063096.t001.
Table 1  Baseline Characteristics of the Study.
Characteristics Traditional prospective ECG gating Doppler-guided prospective ECG gating Total P value
No. of patients 61 58 119
Clinical feature
Typical angina, n (%) 19 (31.2) 23 (39.6) 42 (35.3) 0.33
Atypical angina, n (%) 15 (24.6) 15 (25.9) 30 (25.2) 0.87
Abnormal ECG, n (%) 15 (24.6) 11 (19.0) 26 (21.9) 0.46
MI, n (%) 12 (19.7) 9 (15.5) 21 (17.6) 0.55
Demographics
Age (y) 63.4±10.9 62.0±11.5 62.7±11.2 0.51
Male/female 42/19 43/15 85/34 0.52
BMI (kg/m2) 24.6±3.6 24.7±3.6 24.6±3.6 0.94
HR during scan (bpm) 75.0±7.7 76.5±9.0 75.7±8.4 0.32

Note: ECG = electrocardiogram; MI = myocardial infarction; BMI = body mass index, calculated as weight in kilograms divided by the square of height in meters; HR = heart rate.


[TableWrap ID: pone-0063096-t002] doi: 10.1371/journal.pone.0063096.t002.
Table 2  Image Quality in Different Sub-groups.
Traditional prospective ECG gating Doppler-guided prospective ECG gating
65–80 bpm ≥80 bpm 65–80 bpm ≥80 bpm
NO. 49 12 43 16
Heart rate (bpm) 71.8±4.0 87.7±5.4 71.9±4.1 88.5±7.3
Image quality
Grade 1 85% (605/710) 72% (127/177) 82% (501/608) 78% (186/237)
Grade 2 11% (81/710) 20% (36/177) 13% (78/608) 12% (29/237)
Grade 3 3% (20/710) 7% (12/177) 4% (24/608) 8% (18/237)
Grade 4 1% (4/710) 1% (2/177) 1% (5/608) 2% (4/237)

[TableWrap ID: pone-0063096-t003] doi: 10.1371/journal.pone.0063096.t003.
Table 3  Diagnostic Accuracy of Traditional and Doppler-Guided Group at Patient-, Artery- and Segment- levels.
Sensitivity(95% CI) Specificity(95% CI) PPV(95% CI) NPV(95% CI)
Traditional Doppler Traditional Doppler Traditional Doppler Traditional Doppler
Patient level 97.0(84.7, 99.5) 96.8(83.8, 99.4) 89.3(72.8, 96.3) 88.9(71.9, 96.2) 91.4(77.6, 97.0) 90.1(76.4, 96.9) 96.2(81.1, 99.3) 96.0(80.5, 99.3)
Artery level 91.1(79.3, 96.5) 93.6(82.8, 97.8) 96.5(93.0, 98.3) 97.8(94.6, 99.2) 85.4(72.8, 92.8) 91.7(80.5, 96.7) 98.0(94.9, 99.2) 98.4(95.3, 99.4)
LM 100.0(20.7, 100.0) 100.0(34.2, 100.0) 100.0(94.0, 100.0) 100.0(93.6, 100.0) 100.0(20.7, 100.0) 100.0(34.2, 100.0) 100.0(94.0, 100.0) 100.0(93.6, 100.0)
LAD 93.8(71.7, 98.9) 92.7(68.5, 98.7) 95.6(85.2, 98.8) 97.7(88.2, 99.6) 88.2(65.7, 96.7) 92.9(68.5, 98.7) 97.7(88.2, 99.6) 97.7(88.2, 99.6)
LCX 92.3(66.7, 98.6) 88.9(67.2, 96.9) 93.9(83.5, 97.9) 95.0(83.5, 98.6) 80.0(54.8, 93.0) 88.9(67.2, 96.9) 97.9(88.9, 99.6) 95.0(83.5, 98.6)
RCA 86.7(62.1, 96.3) 100.0(77.2, 100.0) 95.7(85.5, 98.8) 97.8(88.4, 99.6) 86.7(62.1, 96.3) 92.9(68.5, 98.7) 95.7(85.5, 98.8) 100.0(92.0, 100.0)
Segment level 95.5(90.9, 97.8) 94.3(89.5, 97.0) 98.0(96.7, 98.8) 97.1(95.6, 98.2) 90.7(85.3, 94.3) 88.2(82.4, 92.2) 99.0 (98.0, 99.5) 98.7(97.5, 99.3)

Note: PPV = Positive predictive value; NPV = Negative predictive value.



Article Categories:
  • Research Article
Article Categories:
  • Medicine
    • Anatomy and Physiology
      • Cardiovascular System
        • Cardiovascular Anatomy
    • Cardiovascular
      • Cardiovascular Imaging
    • Clinical Research Design
    • Diagnostic Medicine
      • Test Evaluation
    • Radiology
      • Diagnostic Radiology
        • Computed Tomography
      • Interventional Radiology
        • Angiography


Previous Document:  ExtremeDB: A Unified Web Repository of Extremophilic Archaea and Bacteria.
Next Document:  MiR-339-5p Regulates the Growth, Colony Formation and Metastasis of Colorectal Cancer Cells by Targe...