Abstract/OtherAbstract:
|
BACKGROUND: Current continuous glucose monitoring (CGM) systems measure glucose levels in the interstitial fluid to estimate blood glucose concentration. A lag time has been observed between CGM system glucose readings and blood glucose levels when glucose levels are changing. Although this lag has been attributed to the time it takes glucose to equilibrate between blood and interstitial fluid compartments, it is unclear to what extent these inaccuracies reflect an intrinsic delay of the device itself. METHODS: Four Guardian® REAL-Time CGM systems (CGMSs) (Medtronic Diabetes, Minimed, CA) and eight glucose sensors were tested in glucose solutions prepared in Krebs bicarbonate buffers at 37 °C. Glucose readings obtained from CGMSs were compared with actual glucose concentrations during controlled changes in glucose concentration performed at four rates (30, 90, and 220 mg/dl/hr(-1) and an instantaneous change of 110 mg/dl) using a linear gradient maker. RESULTS: Irrespective of the rate and direction of changes in glucose concentration, the readings obtained from CGMSs were significantly different from actual glucose levels. The faster the rise or fall in actual glucose concentration, the more pronounced the mismatch with CGMS glucose readings. Furthermore, the intrinsic lag times (8.3 to 40.1 min) were high enough to account for the lags reported in previous in vivo studies. CONCLUSIONS: The lag intrinsic of the CGMS may make a significant contribution to the mismatch between CGM system readings and blood glucose concentrations. |
| | |
Authors:
|
Raymond J Davey; Chee Low; Timothy W Jones; Paul A Fournier |
Related Documents
:
|
1345575 - Thyrotropin-releasing hormone (trh) does not suppress growth hormone response to l-dopa... 11317665 - Close relation of fasting insulin-like growth factor binding protein-1 (igfbp-1) with g... 7584015 - Regulation of protein metabolism during stress. 17336735 - Effects of growth hormone on glucose and fat metabolism in human subjects. 9058705 - Insulin-like growth factor binding protein-1 is elevated in patients with polycythemia ... 11565495 - Postoperative management of the diabetic patient. |
Publication Detail:
|
Type: Comparative Study; Journal Article; Research Support, Non-U.S. Gov't Date: 2010-11-01 |
Journal Detail:
|
Title: Journal of diabetes science and technology Volume: 4 ISSN: 1932-2968 ISO Abbreviation: J Diabetes Sci Technol Publication Date: 2010 Nov |
Date Detail:
|
Created Date: 2010-12-06 Completed Date: 2011-03-25 Revised Date: 2011-11-01 |
Medline Journal Info:
|
Nlm Unique ID: 101306166 Medline TA: J Diabetes Sci Technol Country: United States |
Other Details:
|
Languages: eng Pagination: 1393-9 Citation Subset: IM |
Copyright Information:
|
© 2010 Diabetes Technology Society. |
Affiliation:
|
School of Sport Science, Exercise and Health, The University of Western Australia, Crawley, Western Australia, Australia. daveyr01@student.uwa.edu.au |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Biological Markers
/
blood Blood Glucose / metabolism* Buffers Equipment Design Extracellular Fluid / metabolism* Humans Kinetics Materials Testing Monitoring, Physiologic* / instrumentation, methods Predictive Value of Tests Reproducibility of Results Temperature |
| Chemical | |
Reg. No./Substance:
|
0/Biological Markers; 0/Blood Glucose; 0/Buffers |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Correlation between plantar foot temperature and diabetic neuropathy: a case study by using an infra...
Next Document: Multicenter user evaluation of ACCU-CHEK® Combo, an integrated system for continuous subcutaneous i...