Document Detail


Continuous-flow cardiac assistance: effects on aortic valve function in a mock loop.
MedLine Citation:
PMID:  20828746     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
BACKGROUND: As the use of left ventricular assist devices (LVADs) to treat end-stage heart failure has become more widespread, leaflet fusion-with resul-tant aortic regurgitation-has been observed more frequently. To quantitatively assess the effects of nonpulsatile flow on aortic valve function, we tested a continuous-flow LVAD in a mock circulatory system (MCS) with an interposed valve.
MATERIALS AND METHODS: To mimic the hemodynamic characteristics of LVAD patients, we utilized an MCS in which a Jarvik 2000 LVAD was positioned at the base of a servomotor-operated piston pump (left ventricular chamber). We operated the LVAD at 8000 to 12,000 rpm, changing the speed in 1000-rpm increments. At each speed, we first varied the outflow resistance at a constant stroke volume, then varied the stroke volume at a constant outflow resistance. We measured the left ventricular pressure, aortic pressure, pump flow, and total flow, and used these values to compute the change, if any, in the aortic duty cycle (aortic valve open time) and transvalvular aortic pressure loads.
RESULTS: Validation of the MCS was demonstrated by the simulation of physiologic pressure and flow waveforms. At increasing LVAD speeds, the mean aortic pressure load steadily increased, while the aortic duty cycle steadily decreased. Changes were consistent for each MCS experimental setting, despite variations in stroke volume and outflow resistance.
CONCLUSIONS: Increased LVAD flow results in an impaired aortic valve-open time due to a pressure overload above the aortic valve. Such an overload may initiate structural changes, causing aortic leaflet fusion and/or regurgitation.
Authors:
Egemen Tuzun; Marcel Rutten; Marco Dat; Frans van de Vosse; Cihan Kadipasaoglu; Bas de Mol
Related Documents :
17767526 - A nomogram for measurement of mitral valve area by proximal isovelocity surface area me...
1868496 - Hemodynamic changes during retrograde left-heart catheterization in patients with aorti...
17690156 - Left atrial remodelling early after mitral valve repair for degenerative mitral regurgi...
8112906 - Pulmonary hemodynamics at rest and effort, 6 and 12 months after mitral valve replaceme...
9163556 - Asymmetrical regulation of blood pressure in the rat medulla.
12366016 - Superconductivity and quantum criticality in cecoin5.
Publication Detail:
Type:  Journal Article     Date:  2010-06-12
Journal Detail:
Title:  The Journal of surgical research     Volume:  171     ISSN:  1095-8673     ISO Abbreviation:  J. Surg. Res.     Publication Date:  2011 Dec 
Date Detail:
Created Date:  2011-11-14     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0376340     Medline TA:  J Surg Res     Country:  United States    
Other Details:
Languages:  eng     Pagination:  443-7     Citation Subset:  IM    
Copyright Information:
Copyright © 2011 Elsevier Inc. All rights reserved.
Affiliation:
Center for Cardiac Support and the Cardiovascular Research Laboratories, Texas Heart Institute at St. Luke's Episcopal Hospital, Houston, Texas; Department of Biomedical Engineering, Eindhoven University of Technology, Eindhoven, The Netherlands; Department of Cardiovascular Surgery, Academic Medical Center, Amsterdam, The Netherlands.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  The Humoral Response After Laparoscopic Versus Open Colorectal Surgery: A Meta-Analysis.
Next Document:  Leupeptin Preserves Cardiac Nitric Oxide Synthase 3 During Reperfusion Following Long-Term Cardiople...