Document Detail


Consideration of kaolinite interference correction for quartz measurements in coal mine dust.
MedLine Citation:
PMID:  23767881     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
Kaolinite interferes with the infrared analysis of quartz. Improper correction can cause over- or underestimation of silica concentration. The standard sampling method for quartz in coal mine dust is size selective, and, since infrared spectrometry is sensitive to particle size, it is intuitively better to use the same size fractions for quantification of quartz and kaolinite. Standard infrared spectrometric methods for quartz measurement in coal mine dust correct interference from the kaolinite, but they do not specify a particle size for the material used for correction. This study compares calibration curves using as-received and respirable size fractions of nine different examples of kaolinite in the different correction methods from the National Institute for Occupational Safety and Health Manual of Analytical Methods (NMAM) 7603 and the Mine Safety and Health Administration (MSHA) P-7. Four kaolinites showed significant differences between calibration curves with as-received and respirable size fractions for NMAM 7603 and seven for MSHA P-7. The quartz mass measured in 48 samples spiked with respirable fraction silica and kaolinite ranged between 0.28 and 23% (NMAM 7603) and 0.18 and 26% (MSHA P-7) of the expected applied mass when the kaolinite interference was corrected with respirable size fraction kaolinite. This is termed "deviation," not bias, because the applied mass is also subject to unknown variance. Generally, the deviations in the spiked samples are larger when corrected with the as-received size fraction of kaolinite than with the respirable size fraction. Results indicate that if a kaolinite correction with reference material of respirable size fraction is applied in current standard methods for quartz measurement in coal mine dust, the quartz result would be somewhat closer to the true exposure, although the actual mass difference would be small. Most kinds of kaolinite can be used for laboratory calibration, but preferably, the size fraction should be the same as the coal dust being collected.
Authors:
Taekhee Lee; William P Chisholm; Michael Kashon; Rosa J Key-Schwartz; Martin Harper
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Journal of occupational and environmental hygiene     Volume:  10     ISSN:  1545-9632     ISO Abbreviation:  J Occup Environ Hyg     Publication Date:  2013 Aug 
Date Detail:
Created Date:  2013-06-17     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  101189458     Medline TA:  J Occup Environ Hyg     Country:  England    
Other Details:
Languages:  eng     Pagination:  425-34     Citation Subset:  IM    
Affiliation:
a Exposure Assessment Branch, National Institute for Occupational Safety and Health , Centers for Disease Control and Prevention , Morgantown , West Virginia.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Synthesis of ?-C-GlcNAc Ser from ?-C-Glc Ser.
Next Document:  Pd-catalyzed cross-coupling of potassium alkenyltrifluoroborates with 2-chloroacetates and 2-chloroa...