Complex flow patterns in a real-size intracranial aneurysm phantom: phase contrast MRI compared with particle image velocimetry and computational fluid dynamics.
MedLine Citation:
PMID:  21480417     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
The aim of this study was to validate the flow patterns measured by high-resolution, time-resolved, three-dimensional phase contrast MRI in a real-size intracranial aneurysm phantom. Retrospectively gated three-dimensional phase contrast MRI was performed in an intracranial aneurysm phantom at a resolution of 0.2 × 0.2 × 0.3 mm(3) in a solenoid rat coil. Both steady and pulsatile flows were applied. The phase contrast MRI measurements were compared with particle image velocimetry measurements and computational fluid dynamics simulations. A quantitative comparison was performed by calculating the differences between the magnitude of the velocity vectors and angles between the velocity vectors in corresponding voxels. Qualitative analysis of the results was executed by visual inspection and comparison of the flow patterns. The root-mean-square errors of the velocity magnitude in the comparison between phase contrast MRI and computational fluid dynamics were 5% and 4% of the maximum phase contrast MRI velocity, and the medians of the angle distribution between corresponding velocity vectors were 16° and 14° for the steady and pulsatile measurements, respectively. In the phase contrast MRI and particle image velocimetry comparison, the root-mean-square errors were 12% and 10% of the maximum phase contrast MRI velocity, and the medians of the angle distribution between corresponding velocity vectors were 19° and 15° for the steady and pulsatile measurements, respectively. Good agreement was found in the qualitative comparison of flow patterns between the phase contrast MRI measurements and both particle image velocimetry measurements and computational fluid dynamics simulations. High-resolution, time-resolved, three-dimensional phase contrast MRI can accurately measure complex flow patterns in an intracranial aneurysm phantom. Copyright © 2011 John Wiley & Sons, Ltd.
Authors:
P Ooij; A Guédon; C Poelma; J Schneiders; M C M Rutten; H A Marquering; C B Majoie; E Vanbavel; A J Nederveen
Related Documents :
22989147 - Utility of 18-fluorodeoxyglucose positron emission/computed tomography in the managemen...
12533337 - Unusual appearance of a cisternal hydatid cyst.
7611087 - Mr demonstration of wegener granulomatosis of the infundibulum, a cause of diabetes ins...
12603527 - Assessment of cancer cell differentiation in small hepatocellular carcinoma by computed...
22634487 - Methanol poisoning: characteristic mri findings.
16159087 - Ct perfusion cerebral blood flow imaging in neurological critical care.
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2011-4-8
Journal Detail:
Title:  NMR in biomedicine     Volume:  -     ISSN:  1099-1492     ISO Abbreviation:  -     Publication Date:  2011 Apr 
Date Detail:
Created Date:  2011-4-11     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8915233     Medline TA:  NMR Biomed     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Copyright Information:
Copyright © 2011 John Wiley & Sons, Ltd.
Affiliation:
Department of Radiology, Academic Medical Center, University of Amsterdam, Amsterdam, the Netherlands; Department of Biomedical Engineering and Physics, Academic Medical Center, University of Amsterdam, Amsterdam, the Netherlands. pim.vanooij@gmail.com.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Effects of Astaxanthin on Oxidative Stress in Overweight and Obese Adults.
Next Document:  Azide-based Cross-Linking of Polymers of Intrinsic Microporosity (PIMs) for Condensable Gas Separati...