Document Detail


Complete transcatheter closure of a patent arterial duct with subsequent haemolysis.
MedLine Citation:
PMID:  20334722     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
We report a development of severe haemolysis after complete transcatheter closure of patent arterial duct. Aortography and echocardiography revealed no signs of residual shunt. Haemolysis occurred a day after the implantation. Aortography was performed and the extrusion of coil in aorta was evident. The extruded part of the coil was surgically removed. No signs of haemolysis remained.
Authors:
Neven Cace; Vladimir Ahel; Iva Bilic
Publication Detail:
Type:  Journal Article     Date:  2010-03-25
Journal Detail:
Title:  Cardiology in the young     Volume:  20     ISSN:  1467-1107     ISO Abbreviation:  Cardiol Young     Publication Date:  2010 Aug 
Date Detail:
Created Date:  2010-06-25     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9200019     Medline TA:  Cardiol Young     Country:  England    
Other Details:
Languages:  eng     Pagination:  462-4     Citation Subset:  IM    
Affiliation:
Department of Pediatrics, University Hospital Center Rijeka, Rijeka, Croatia. ncace1@gmail.com