Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use.
MedLine Citation:
PMID:  23263536     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
OBJECTIVE(S): We compared a convenient, cuffless, wrist-type ambulatory blood pressure monitoring (ABPM) device (BPro) and a standard arm-type ABPM monitor under ambulatory conditions. METHODS: Fifty normotensive and prehypertensive volunteers who had no interarm differences in BP were enrolled. The wrist and arm monitors were attached to the left wrist and the right arm, respectively, and provided readouts at 15 and 30-min intervals, respectively. Ambulatory BP levels and the extent of agreement by intraclass correlation were evaluated. In 15 of the 50 participants, we also examined values at different arm positions: heart level, above the head, and hanging at the sides. RESULTS: The awake mean systolic blood pressure (SBP) values (122±13 vs. 127±11 mmHg, P<0.01) were significantly lower with the wrist monitor than the arm monitor, and the mean sleep diastolic blood pressure (DBP) (71±8 vs. 64±8 mmHg, P<0.01) was significantly higher with the wrist monitor than the arm monitor. The intraclass correlation values between the monitors were 0.54 for 24 h SBP and 0.52 for awake SBP, considered to indicate a moderate agreement. The BP values in the arm-raised position were significantly higher for the wrist monitor than for the arm monitor (SBP: 129±14 vs. 108±14 mmHg, P<0.01; DBP: 83±13 vs. 64±11 mmHg, P<0.01). However, the SBP and DBP values in the other arm positions were similar between the monitors. CONCLUSION: The wrist monitor showed fair agreement with the arm monitor in the ambulatory condition, and was stable irrespective of arm positions.
Authors:
Takahiro Komori; Kazuo Eguchi; Satoshi Hoshide; Bryan Williams; Kazuomi Kario
Related Documents :
21884956 - Decreasing sleep-time blood pressure determined by ambulatory monitoring reduces cardio...
22154796 - Power loss and right ventricular efficiency in patients after tetralogy of fallot repai...
7213716 - The size of erythrocyte ghosts.
21956526 - Prognostic significance of home arterial stiffness index derived from self-measurement ...
22781266 - Side-hole two-core microstructured optical fiber for hydrostatic pressure sensing.
7708716 - Genetic control of blood pressure and the angiotensinogen locus.
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2012-12-20
Journal Detail:
Title:  Blood pressure monitoring     Volume:  -     ISSN:  1473-5725     ISO Abbreviation:  Blood Press Monit     Publication Date:  2012 Dec 
Date Detail:
Created Date:  2012-12-24     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
aDepartment of Medicine, Division of Cardiovascular Medicine, Jichi Medical University, Shimotsuke, Tochigi, Japan bInstitute of Cardiovascular Science, University College London, London, UK.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Cost-effectiveness of secondary screening modalities for hypertension.
Next Document:  Asleep home blood pressure monitoring in obstructive sleep apnea: a pilot study.