Comparison of various agents in contrast enhancement of color Doppler flow images: an in vitro study.
MedLine Citation:
PMID:  8456528     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The commercially manufactured contrast agents, Echovist and Albunex, were compared with sonicated conventional agents, indocyanine green, 29% renografin-60, 0.9% normal saline and 25% mannitol in their ability to enhance color Doppler flow signals. In a pneumatically regulated pulsatile flow system, a glycerine, saline (0.9%) and sand (5 microns particle size) solution was imaged using a 2.5 MHz phased-array transducer. Four different flow velocities (0.40, 0.35, 0.30 and 0.25 m/s) as measured by color Doppler guided pulse Doppler were utilized. All color Doppler settings were kept constant throughout the study. Utilizing a power injector, four different volumes (1.0, 1.5, 2.0 and 2.5 mL) of each contrast agent were injected into the flow medium at various transducer angles (20, 30, 40 and 60 degrees) and various distances from the transducer (3.38, 5.5, 6.76 cm). For Echovist and Albunex, several concentrations varying from 2% to 100% were used. Keeping instrument settings constant, color flow areas obtained before and after each contrast injection were planimeterized and the percent increase in the color flow area computed and compared. At full (100%) concentration, 20 degrees transducer angle and a flow velocity of 0.40 m/s, the maximum increase in the color flow area was 568%, 251%, 180%, 110%, 71%, and 38% for Echovist, Albunex, sonicated indocyanine green, renografin, normal saline and mannitol, respectively. However, a significant reduction in the degree of color flow enhancement was observed, with decreases in the concentration of these agents, and increases in the Doppler beam incident angle or distance from the transducer. Increasing the flow velocity of the medium into which contrast was injected did not produce significant changes in the contrast enhancement effect for all agents except Echovist. Increasing the injection volume significantly increased the color flow area for sonicated agents but not for Echovist or Albunex. This preliminary in vitro study shows that the commercially manufactured contrast agents, Echovist and Albunex, are much superior to sonicated conventional contrast agents in the enhancement of color Doppler flow signals. Of the sonicated agents, indocyanine green had the best enhancement capability.
Authors:
P Fan; P J Czuwala; N C Nanda; S M Rosenthal; A Yoganathan
Related Documents :
8065128 - A rapid spectrophotometric blood cyanide determination applicable to emergency toxicology.
3389968 - Effect of kallikrein on testicular blood circulation.
16356088 - Measurement of thyroid blood flow area is useful for diagnosing the cause of thyrotoxic...
7648958 - An objective measure of stool color for differentiating upper from lower gastrointestin...
6732508 - Stereoselectivity of enzymes involved in toxicity and detoxification of soman.
10696708 - Flow dependence and time constant of the change in nitric oxide concentration measured ...
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  Ultrasound in medicine & biology     Volume:  19     ISSN:  0301-5629     ISO Abbreviation:  Ultrasound Med Biol     Publication Date:  1993  
Date Detail:
Created Date:  1993-04-20     Completed Date:  1993-04-20     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  0410553     Medline TA:  Ultrasound Med Biol     Country:  ENGLAND    
Other Details:
Languages:  eng     Pagination:  45-57     Citation Subset:  IM    
Affiliation:
Division of Cardiovascular Disease, University of Alabama, Birmingham 35233.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Blood Flow Velocity
Color
Contrast Media*
Humans
Image Enhancement / methods*
Models, Biological
Observer Variation
Pulsatile Flow*
Reproducibility of Results
Sensitivity and Specificity
Ultrasonography / methods*
Chemical
Reg. No./Substance:
0/Contrast Media

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Effect of perfusion and blood content on ultrasonic backscattering of liver tissue.
Next Document:  Thickness of tissues intervening between the transducer and fetus and models for fetal exposure calc...