Comparison of two point-of-care testing (POCT) devices for fetal lactate during labor.
MedLine Citation:
PMID:  21955187     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
Abstract Background: Point-of-care testing (POCT) of fetal scalp blood lactate is used as an alternative for pH analysis. Lactate measurements have not been standardized and values vary with each device used. The aim of this study was to evaluate the performance of two POCT lactate meters for intrapartum use. Methods: Analytical performance of StatStrip Lactate (Nova Biomedical) and Lactate Pro (Arkray) was evaluated using CLSI EP10. Both POCT meters were compared with our lactate reference method (RapidLab 860; Siemens Healthcare Diagnostics) using fetal scalp and neonatal cord blood. Deming regression analysis was performed. Results: StatStrip Lactate coefficients of variation (CVs) were 5.1%, 5.0% and 2.6% at 0.9, 7.5 and 14.1 mmol/L lactate, respectively. CVs for Lactate Pro were 10.7%, 5.2% and 5.7% at 1.7, 4.1 and 6.4 mmol/L lactate, respectively. Consecutive lactate measurements in 37 fetal scalp and 122 cord blood samples revealed different test characteristics for the two POCT devices. In fetal scalp blood: StatStrip Lactate=1.13*RapidLab-0.39 (R(2)=0.907) and Lactate Pro=0.95*RapidLab-0.03 (R(2)=0.823). In cord artery blood: StatStrip Lactate=1.08*RapidLab-0.09 (R(2)=0.810) and Lactate Pro=0.72*RapidLab+0.59 (R(2)=0.807). Conclusions: Overall performance of both Lactate Pro and StatStrip Lactate was good, with StatStrip Lactate having smallest CVs and closest correlation to our reference method. Both StatStrip Lactate and Lactate Pro can be used as a lactate POCT device for obstetric use.
Authors:
Ayesha M F Heinis; Jacqueline Dinnissen; Marc E A Spaanderman; Fred K Lotgering; Jacqueline M T Klein Gunnewiek
Related Documents :
6298497 - Flow-correlated influx of k, ca, p, and mg during continuous microperfusion of the loop...
15218897 - Oral health condition and saliva flow in southern chinese with sjögren's syndrome.
8884927 - Mathematical modelling of the concentration of microorganisms in the urinary bladder un...
19223027 - Continuous analysis of parotid saliva during resting and short-duration simulated chewing.
17756837 - Somali current: evolution of surface flow.
3780557 - Redistribution of ovarian blood flow after injection of human chorionic gonadotropin an...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2011-9-29
Journal Detail:
Title:  Clinical chemistry and laboratory medicine : CCLM / FESCC     Volume:  -     ISSN:  1434-6621     ISO Abbreviation:  -     Publication Date:  2011 Sep 
Date Detail:
Created Date:  2011-9-29     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9806306     Medline TA:  Clin Chem Lab Med     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
Department of Obstetrics and Gynecology, Radboud University Nijmegen Medical Center, Nijmegen, The Netherlands.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Diurnal variability of total calcium during normal sleep and after an acute shift of sleep.
Next Document:  Use of memantine for the treatment of dementia.