Comparison of two flow generators with a noninvasive ventilator to deliver continuous positive airway pressure: a test lung study.
MedLine Citation:
PMID:  16222519     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: To compare the performance of two continuous flow generators with a ventilator designed for noninvasive positive pressure ventilation (NPPV) to deliver continuous positive airway pressure (CPAP). The performance of flow generators using different oxygen pressure supplies was also compared. DESIGN AND SETTING: Experimental study using a mechanical lung model in a university research laboratory. MEASUREMENTS: Two flow generators supplied at 100, 200, and 300 kPa and an NPPV ventilator were compared at CPAP of 5, 10, and 15 cmH2O in: (a) area under the adjusted CPAP level during inspiration, (b) capacity to attain the preset CPAP, and (c) tidal volume. RESULTS: The NPPV ventilator attained the preset CPAP better than flow generators, but its area under adjusted CPAP was similar to or higher than that of flow generators when these were adjusted to their better pressure supply. Both flow generators had better performance with an output flow around 100 l/min, which was achieved at 100 kPa with one flow generator and 300 with the other. Flow generators and the NPPV ventilator generated similar tidal volumes. CONCLUSIONS: Flow generators performance showed large variations among different devices and oxygen pressure supplies. Adjusted to their better pressure supply, flow generators had a similar or better capacity to maintain the CPAP level, but the NPPV ventilator was more reliable to attain the preset CPAP. Flow generators could be an alternative to provide CPAP in low-income areas, usually with scarce medical equipment availability.
Authors:
Carolina Fu; Pedro Caruso; Jeanette Janaina Jaber Lucatto; Guilherme Pinto de Paula Schettino; Rogério de Souza; Carlos Roberto Ribeiro Carvalho
Related Documents :
2399779 - Monitoring airway pressure: a new standard.
3063709 - Parenchymal interdependence and airway response to methacholine in excised dog lobes.
2132339 - Predicting arterial oxygenation during one-lung ventilation with continuous positive ai...
15994269 - Sleep x 9: an approach to treatment of obstructive sleep apnoea/hypopnoea syndrome incl...
11982839 - Higher airway seal does not imply adequate positioning of laryngeal mask airways in pae...
9366939 - Laryngeal mask bite blocks--rolled gauze versus guedel airway.
21383619 - Adaptive support ventilation may deliver unwanted respiratory rate-tidal volume combina...
14625119 - Tracheal epinephrine or norepinephrine preceded by beta blockade in a dog model. can be...
18529369 - Electro-active polymer transduction for distributed netted sensing.
Publication Detail:
Type:  Comparative Study; Journal Article     Date:  2005-10-13
Journal Detail:
Title:  Intensive care medicine     Volume:  31     ISSN:  0342-4642     ISO Abbreviation:  Intensive Care Med     Publication Date:  2005 Nov 
Date Detail:
Created Date:  2005-11-11     Completed Date:  2006-01-31     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  7704851     Medline TA:  Intensive Care Med     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1587-91     Citation Subset:  IM    
Affiliation:
Physiotherapy, Phonology and Occupational Therapy Department, Medical School, University of São Paulo, São Paulo, Brazil.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Artificial Organs
Continuous Positive Airway Pressure / instrumentation*
Lung*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine