Comparison of topcon optical coherence tomography and ultrasound pachymetry.
MedLine Citation:
PMID:  23190715     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
PURPOSE: To establish within-rater repeatability and minimum measurements required for reliable assessment of central corneal thickness (CCT) using Topcon three-dimensional OCT-2000 (Topcon Medical Systems, Oakland, NJ) Fourier-domain optical coherence tomography (FD-OCT) and to test agreement against ultrasound pachymetry (USP).
METHODS: Twenty participants underwent five scans using FD-OCT followed by five CCT measurements using USP. Each FD-OCT scan produced 12 meridional optical sections, giving 60 sections in total. The FD-OCT CCT was calculated from each section using Topcon FastMap software. Within-rater repeatability was assessed using intraclass correlation coefficient (ICC), within-subject variance (Sw), and within-subject coefficients of repeatability (COR) and variation (COV). Agreement was assessed using analysis of variance, simple linear regression, Bland-Altman analysis, and ICC. The minimum number of FD-OCT scans required to achieve a mean CCT within the 99% confidence intervals of all 60 measurements was calculated.
RESULTS: Mean CCTs were 528 ± 27 μm (FD-OCT) and 544 ± 29 μm (USP). The FD-OCT within-rater repeatability values assessed using ICC, Sw, COR, and COV were 0.99, 5.07 μm, 14.07 μm, and 0.01%, respectively. The USP within-rater repeatability values assessed using ICC, Sw, COR, and COV were 0.96, 5.49 μm, 15.22 μm, and 0.01%, respectively. Both instruments were similarly correlated (p < 0.05), with FD-OCT underestimating CCT by 16.08 μm relative to USP. Four FD-OCT scans were required to achieve a mean within the 99% confidence interval of all measurements for each participant.
CONCLUSIONS: Both instruments displayed high repeatability, with strong agreement between devices, although FD-OCT significantly underestimated CCT compared with USP. The FD-OCT agreement with USP remained strong when only four scans were analyzed.
Authors:
Luke Cameron Northey; Paul Gifford; Gavin C Boneham
Related Documents :
21405635 - Entropic stabilization of tunable planar modulated superstructures.
21230545 - Dynamics-based centrality for directed networks.
21361585 - Transmission line based short pulse generation circuits in a 0.13 μm complementary met...
21230595 - Structures in magnetohydrodynamic turbulence: detection and scaling.
22540585 - Generation of high-energy few-cycle laser pulses by using the ionization-induced self-c...
17415385 - Analysis of multimask fabrication errors for diffractive optical elements.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Optometry and vision science : official publication of the American Academy of Optometry     Volume:  89     ISSN:  1538-9235     ISO Abbreviation:  Optom Vis Sci     Publication Date:  2012 Dec 
Date Detail:
Created Date:  2012-11-29     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8904931     Medline TA:  Optom Vis Sci     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1708-14     Citation Subset:  IM    
Affiliation:
*BSc(adv) †PhD, FAAO ‡PhD School of Medicine Sydney, The University of Notre Dame Australia, Broadway (LCN); and School of Optometry and Vision Science, The University of New South Wales, Sydney (PG, GCB), New South Wales, Australia.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Visual Strehl Performance of IOL Designs with Extended Depth of Focus.
Next Document:  Concurrent vision dysfunctions in convergence insufficiency with traumatic brain injury.